A Validated Reversed-Phase HPLC Method for the Determination of Atorvastatin Calcium in Tablets

A Reversed-Phase Liquid Chromatographic (RP-LC) assay method was developed for the quantitative determination of atorvastatin calcium in the presence of its degradation products. The assay involved an isocratic elution of atorvastatin calcium in a LiChroCARTR 250*4 mm HPLC Cartridge LiChrospherR 100...

Full description

Bibliographic Details
Authors: Simionato, Laura Daniela, Ferello, L., Stamer. S., Repetto, M. F., Zubata, P. D., Segall, Adriana Ines
Format: article
Status:Published version
Publication Date:2014
Country:Argentina
Institution:Consejo Nacional de Investigaciones Científicas y Técnicas
Repository:CONICET Digital (CONICET)
Language:English
OAI Identifier:oai:ri.conicet.gov.ar:11336/30538
Online Access:http://hdl.handle.net/11336/30538
Access Level:Open access
Keyword:Atorvastatin calcium
RP-HPLC
Tablets assay
https://purl.org/becyt/ford/1.4
https://purl.org/becyt/ford/1
Description
Summary:A Reversed-Phase Liquid Chromatographic (RP-LC) assay method was developed for the quantitative determination of atorvastatin calcium in the presence of its degradation products. The assay involved an isocratic elution of atorvastatin calcium in a LiChroCARTR 250*4 mm HPLC Cartridge LiChrospherR 100 RP-18 (5 μm) column using a mobile phase consisting of 0.1% acetic acid solution: acetonitrile (45:55, v/v), pH = 3.8. The flow rate was 0.8 mL/min and the analytes monitored at 246 nm. The assay method was found to be linear from 8.13 to 23.77 μg/mL. All the validation parameters were within the acceptance range. The developed method was successfully applied to estimate the amount of atorvastatin calcium in tablets.