Resolution of a clinical AmpliChip CYP450 Test (TM) no call: discovery and characterization of novel CYP2D6*1 haplotypes

A Han Chinese patient failed CYP2D6 genotype analysis with the AmpliChip CYP450 Test (TM). The CYP2D6 gene locus of the patient and her son were extensively genotyped including copy number variation and gene resequencing. Two SNPs were discovered on the patient's CYP2D6*1 allele, -498C>A and...

Descripción completa

Detalles Bibliográficos
Autores: Gaedigk, A, Garcia-Ribera, C, Jeong, HE, Shin, JG, Hernandez-Sanchez, JT
Tipo de recurso: artículo
Estado:Versión publicada
Fecha de publicación:2014
País:España
Institución:Institut d’Investigació Biomèdica Sant Pau (IIB Sant Pau)
Repositorio:r-IIB SANT PAU. Repositorio Institucional de Producción Científica del Instituto de Investigación Biomédica Sant Pau
OAI Identifier:oai:iibsantpau.fundanetsuite.com:p14743
Acceso en línea:https://iibsantpau.fundanetsuite.com/Publicaciones/ProdCientif/PublicacionFrw.aspx?id=14743
Access Level:acceso abierto
Palabra clave:AmpliChip CYP450 Test (TM)
cytochrome CYP450
CYP2D6
CYP2D6*10
genotyping
pharmacogenetics
Descripción
Sumario:A Han Chinese patient failed CYP2D6 genotype analysis with the AmpliChip CYP450 Test (TM). The CYP2D6 gene locus of the patient and her son were extensively genotyped including copy number variation and gene resequencing. Two SNPs were discovered on the patient's CYP2D6*1 allele, -498C>A and 1661G>C, while the son's CYP2D6*1 allele had -498C>A only. AmpliChip failure was attributed to the presence of a CYP2D6*1 allele carrying the 1661G>C SNP. Functional analyses of -498C>A did not reveal altered activity in vitro or in vivo suggesting that both novel CYP2D6*1 subvariants are functional. The implementation of pharmacogenetics-guided drug therapy relies on accurate clinical-grade genotype analysis. Although the AmpliChip is a reliable platform, numerous allelic (sub) variants and gene arrangements are not detected or may trigger no calls. While such cases may be rare, the clinical/genetic testing community must be aware of the challenges of CYP2D6 testing on the AmpliChip platform and implications regarding accuracy of test results.