Analytical parameters of a novel glucose biosensor based on grafted PFM as a covalent immobilization technique

Bioanalytical methods, in particular electrochemical biosensors, are increasingly used in different industrial sectors due to their simplicity, low cost, and fast response. However, to be able to reliably use this type of device, it is necessary to undertake in-depth evaluation of their fundamental...

Descripción completa

Detalles Bibliográficos
Autores: Artigues Cladera, Margalida, Gilabert-Porres, Joan, Texidó Bartes, Robert, Borrós, Salvador, Abella, Jordi, Colominas Fuster, Sergi
Tipo de recurso: artículo
Fecha de publicación:2021
País:España
Institución:Varias* (Consorci de Biblioteques Universitáries de Catalunya, Centre de Serveis Científics i Acadèmics de Catalunya)
Repositorio:Recercat. Dipósit de la Recerca de Catalunya
OAI Identifier:oai:recercat.cat:20.500.14342/3944
Acceso en línea:http://hdl.handle.net/20.500.14342/3944
https://doi.org/10.3390/s21124185
Access Level:acceso abierto
Palabra clave:Bioanalytical methods
Electrochemical biosensor
Hydroxyethyl methacrylate (HEMA)
Pentafluorophenyl methacrylate (PFM)
Glucose oxidase (GOx)
6
Descripción
Sumario:Bioanalytical methods, in particular electrochemical biosensors, are increasingly used in different industrial sectors due to their simplicity, low cost, and fast response. However, to be able to reliably use this type of device, it is necessary to undertake in-depth evaluation of their fundamental analytical parameters. In this work, analytical parameters of an amperometric biosensor based on covalent immobilization of glucose oxidase (GOx) were evaluated. GOx was immobilized using plasma-grafted pentafluorophenyl methacrylate (pgPFM) as an anchor onto a tailored HEMA-co-EGDA hydrogel that coats a titanium dioxide nanotubes array (TiO2NTAs). Finally, chitosan was used to protect the enzyme molecules. The biosensor offered outstanding analytical parameters: repeatability (RSD = 1.7%), reproducibility (RSD = 1.3%), accuracy (deviation = 4.8%), and robustness (RSD = 2.4%). In addition, the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor showed good long-term stability; after 20 days, it retained 89% of its initial sensitivity. Finally, glucose concentrations of different food samples were measured and compared using an official standard method (HPLC). Deviation was lower than 10% in all measured samples. Therefore, the developed biosensor can be considered to be a reliable analytical tool for quantification measurements.