Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia?
BACKGROUND: Correction of exercise hypoxemia in advanced lung diseases is crucial and often challenging. However, oxygen-conserving devices have been introduced in the market with limited evidence of effectiveness. In the present study the efficacy of 2 oxygen-conserving devices, a pulse demand oxyg...
| Autores: | , , , , , , |
|---|---|
| Tipo de recurso: | artículo |
| Estado: | Versión publicada |
| Fecha de publicación: | 2013 |
| País: | España |
| Institución: | Institut d’Investigació Biomèdica Sant Pau (IIB Sant Pau) |
| Repositorio: | r-IIB SANT PAU. Repositorio Institucional de Producción Científica del Instituto de Investigación Biomédica Sant Pau |
| OAI Identifier: | oai:iibsantpau.fundanetsuite.com:p9655 |
| Acceso en línea: | https://iibsantpau.fundanetsuite.com/Publicaciones/ProdCientif/PublicacionFrw.aspx?id=9655 |
| Access Level: | acceso abierto |
| Palabra clave: | COPD conservers exercise test interstitial lung disease oxygen inhalation therapy technology assessment |
| id |
ES_91e5a36e3cf647cf911dd577e100a06e |
|---|---|
| oai_identifier_str |
oai:iibsantpau.fundanetsuite.com:p9655 |
| network_acronym_str |
ES |
| network_name_str |
España |
| repository_id_str |
|
| spelling |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia?Marti, SPajares, VMorante, FRamon, MALara, JFerrer, JGuell, MRCOPDconserversexercise testinterstitial lung diseaseoxygen inhalation therapytechnology assessmentBACKGROUND: Correction of exercise hypoxemia in advanced lung diseases is crucial and often challenging. However, oxygen-conserving devices have been introduced in the market with limited evidence of effectiveness. In the present study the efficacy of 2 oxygen-conserving devices, a pulse demand oxygen delivery (DOD) system and pendant reservoir cannula (PRC), were evaluated in subjects with COPD and interstitial lung disease (ILD). METHODS: A cross-sectional, crossover study included 28 COPD and 31 ILD subjects with oxygen desaturation on the 6-min walk test (average S-pO2 < 88%). Each subject underwent 3 walk tests with DOD, PRC, and continuous oxygen flow by standard nasal cannula (CFNC), in random order, taking average S-pO2 >= 90% as the resaturation criterion. RESULTS: Exercise desaturation was corrected in 79%, 79%, and 86% of COPD subjects with CFNC, DOD, and PRC, respectively, and in 77%, 61%, and 81% of ILD subjects with CFNC, DOD, and PRC, respectively. When compared to CFNC, the oxygen-conserving devices showed similar efficacy, except a lower performance for the DOD in the ILD subjects (P = .01). CONCLUSIONS: Although these oxygen-conserving devices corrected exercise hypoxemia in most COPD and ILD subjects, correction was not achieved in about 20% of the severe COPD subjects, regardless of the device, and in nearly 40% of the ILD subjects with the DOD device. These findings underscore that individualized adjustment of oxygen flow is needed for optimal correction of exercise hypoxemia, especially with a DOD in an ILD patient.DAEDALUS ENTERPRISES INC2013info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttps://iibsantpau.fundanetsuite.com/Publicaciones/ProdCientif/PublicacionFrw.aspx?id=9655Respiratory CareISSN: 00201324ISSNe: 19433654reponame:r-IIB SANT PAU. Repositorio Institucional de Producción Científica del Instituto de Investigación Biomédica Sant Pauinstname:Institut d’Investigació Biomèdica Sant Pau (IIB Sant Pau)Inglésinfo:eu-repo/semantics/openAccessoai:iibsantpau.fundanetsuite.com:p96552026-06-14T12:41:47Z |
| dc.title.none.fl_str_mv |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia? |
| title |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia? |
| spellingShingle |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia? Marti, S COPD conservers exercise test interstitial lung disease oxygen inhalation therapy technology assessment |
| title_short |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia? |
| title_full |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia? |
| title_fullStr |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia? |
| title_full_unstemmed |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia? |
| title_sort |
Are Oxygen-Conserving Devices Effective for Correcting Exercise Hypoxemia? |
| dc.creator.none.fl_str_mv |
Marti, S Pajares, V Morante, F Ramon, MA Lara, J Ferrer, J Guell, MR |
| author |
Marti, S |
| author_facet |
Marti, S Pajares, V Morante, F Ramon, MA Lara, J Ferrer, J Guell, MR |
| author_role |
author |
| author2 |
Pajares, V Morante, F Ramon, MA Lara, J Ferrer, J Guell, MR |
| author2_role |
author author author author author author |
| dc.subject.none.fl_str_mv |
COPD conservers exercise test interstitial lung disease oxygen inhalation therapy technology assessment |
| topic |
COPD conservers exercise test interstitial lung disease oxygen inhalation therapy technology assessment |
| description |
BACKGROUND: Correction of exercise hypoxemia in advanced lung diseases is crucial and often challenging. However, oxygen-conserving devices have been introduced in the market with limited evidence of effectiveness. In the present study the efficacy of 2 oxygen-conserving devices, a pulse demand oxygen delivery (DOD) system and pendant reservoir cannula (PRC), were evaluated in subjects with COPD and interstitial lung disease (ILD). METHODS: A cross-sectional, crossover study included 28 COPD and 31 ILD subjects with oxygen desaturation on the 6-min walk test (average S-pO2 < 88%). Each subject underwent 3 walk tests with DOD, PRC, and continuous oxygen flow by standard nasal cannula (CFNC), in random order, taking average S-pO2 >= 90% as the resaturation criterion. RESULTS: Exercise desaturation was corrected in 79%, 79%, and 86% of COPD subjects with CFNC, DOD, and PRC, respectively, and in 77%, 61%, and 81% of ILD subjects with CFNC, DOD, and PRC, respectively. When compared to CFNC, the oxygen-conserving devices showed similar efficacy, except a lower performance for the DOD in the ILD subjects (P = .01). CONCLUSIONS: Although these oxygen-conserving devices corrected exercise hypoxemia in most COPD and ILD subjects, correction was not achieved in about 20% of the severe COPD subjects, regardless of the device, and in nearly 40% of the ILD subjects with the DOD device. These findings underscore that individualized adjustment of oxygen flow is needed for optimal correction of exercise hypoxemia, especially with a DOD in an ILD patient. |
| publishDate |
2013 |
| dc.date.none.fl_str_mv |
2013 |
| dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion |
| format |
article |
| status_str |
publishedVersion |
| dc.identifier.none.fl_str_mv |
https://iibsantpau.fundanetsuite.com/Publicaciones/ProdCientif/PublicacionFrw.aspx?id=9655 |
| url |
https://iibsantpau.fundanetsuite.com/Publicaciones/ProdCientif/PublicacionFrw.aspx?id=9655 |
| dc.language.none.fl_str_mv |
Inglés |
| language_invalid_str_mv |
Inglés |
| dc.rights.none.fl_str_mv |
info:eu-repo/semantics/openAccess |
| eu_rights_str_mv |
openAccess |
| dc.publisher.none.fl_str_mv |
DAEDALUS ENTERPRISES INC |
| publisher.none.fl_str_mv |
DAEDALUS ENTERPRISES INC |
| dc.source.none.fl_str_mv |
Respiratory Care ISSN: 00201324 ISSNe: 19433654 reponame:r-IIB SANT PAU. Repositorio Institucional de Producción Científica del Instituto de Investigación Biomédica Sant Pau instname:Institut d’Investigació Biomèdica Sant Pau (IIB Sant Pau) |
| instname_str |
Institut d’Investigació Biomèdica Sant Pau (IIB Sant Pau) |
| reponame_str |
r-IIB SANT PAU. Repositorio Institucional de Producción Científica del Instituto de Investigación Biomédica Sant Pau |
| collection |
r-IIB SANT PAU. Repositorio Institucional de Producción Científica del Instituto de Investigación Biomédica Sant Pau |
| repository.name.fl_str_mv |
|
| repository.mail.fl_str_mv |
|
| _version_ |
1869413427330416640 |
| score |
15,812455 |