Real-Time Continuous Glucose Monitoring in an Intensive Care Unit: Better Accuracy in Patients with Septic Shock

This study assessed the accuracy of real-time continuous glucose monitoring system (RTCGMS) devices in an intensive care unit (ICU) to determine whether the septic status of the patient has any influence on the accuracy of the RTCGMS. Subjects and methods: In total, 41 patients on insulin therapy we...

Descripción completa

Detalles Bibliográficos
Autores: Lorencio Cárdenas, Carolina, Leal Moncada, Yenny Teresa, Bonet, Alfonso, Bondia, Jorge, Palerm, César C., Tache, Abdo, Sirvent Calvera, Josep Maria, Vehí, Josep
Tipo de recurso: artículo
Estado:Versión publicada
Fecha de publicación:2012
País:España
Institución:Varias* (Consorci de Biblioteques Universitáries de Catalunya, Centre de Serveis Científics i Acadèmics de Catalunya)
Repositorio:Recercat. Dipósit de la Recerca de Catalunya
OAI Identifier:oai:recercat.cat:10256/8598
Acceso en línea:http://hdl.handle.net/10256/8598
Access Level:acceso abierto
Palabra clave:Glucosa
Glucose
Xoc sèptic
Septic shock
Septicèmia
Septicemia
Descripción
Sumario:This study assessed the accuracy of real-time continuous glucose monitoring system (RTCGMS) devices in an intensive care unit (ICU) to determine whether the septic status of the patient has any influence on the accuracy of the RTCGMS. Subjects and methods: In total, 41 patients on insulin therapy were included. Patients were monitored for 72 h using RTCGMS. Arterial blood glucose (ABG) samples were obtained following the protocol established in the ICU. The results were evaluated using paired values (excluding those used for calibration) with the performance assessed using numerical accuracy. Nonparametric tests were used to determine statistically significant differences in accuracy. Results: In total, 956 ABG/RTCGMS pairs were analyzed. The overall median relative absolute difference (RAD) was 13.5%, and the International Organization for Standardization (ISO) criteria were 68.1%. The median RADs reported for patients with septic shock, with sepsis, and without sepsis were 11.2%, 14.3%, and 16.3%, respectively (P<0.05). Measurements meeting the ISO criteria were 74.5%, 65.6%, and 63.7% for patients with septic shock, with sepsis, and without sepsis, respectively (P<0.05). Conclusions: The results showed that the septic status of patients influenced the accuracy of the RTCGMS in the ICU. Accuracy was significantly better in patients with septic shock in comparison with the other patient cohorts