Detector de luminescência de descargas eletrocirúrgicas

The electrosurgical procedure, used in large surgeries, promotes cut and/or coagulation of the organic tissue. This procedure is possible because the electrical discharge between a metal electrode and the organic tissue. Characteristics of the spark are still almost unknown. This work intends to col...

Full description

Bibliographic Details
Author: Góes, Sandro Waldir
Format: master thesis
Status:Published version
Publication Date:2016
Country:Brasil
Institution:Universidade Tecnológica Federal do Paraná (UTFPR)
Repository:Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT))
Language:Portuguese
OAI Identifier:oai:repositorio.utfpr.edu.br:1/2508
Online Access:http://repositorio.utfpr.edu.br/jspui/handle/1/2508
Access Level:Open access
Keyword:Luminescência
Descargas elétricas
Detectores ópticos
Faiscas elétricas
Engenharia biomédica
Luminescence
Electric discharges
Optical detectors
Electric spark
Biomedical engineering
CNPQ::ENGENHARIAS::ENGENHARIA BIOMEDICA::BIOENGENHARIA
Engenharia Biomédica
Description
Summary:The electrosurgical procedure, used in large surgeries, promotes cut and/or coagulation of the organic tissue. This procedure is possible because the electrical discharge between a metal electrode and the organic tissue. Characteristics of the spark are still almost unknown. This work intends to collaborate to the electrosurgical sparks study, reporting the development of a photodetector device, capable to detect flashes of light in hundreds of kHz, and introducing the waveform of luminescence from the electrosurgical discharge. Methods: The photodetector deviceis composed mainly of a photodiode PIN and an transimpedance amplifier. Spacial care has taken to isolate the device from electromagnetic noise, employing a shield and an optical fiber cable. To obtain, analyze and store data, a specific hardware, software and techniques, have been used and described in detail. Results: From the luminescence signal detected by photodetector device and voltage and current shapes stored in an oscilloscope, it is possible to see the relation between sparks luminescence, voltage and current on positive and negative cycles of the electrical discharge. The luminescence collaborates to show the asymmetry on voltage signals already reported by other researchers. Conclusion: Effective for the study of luminescence emitted by electrosurgical discharges, enabling a new study of the same channel, particularly the speed of response.