A coordinated control strategy for insulin and glucagon delivery in type 1 diabetes

[EN] Type 1 diabetes is an autoimmune condition characterised by a pancreatic insulin secretion deficit, resulting in high blood glucose concentrations, which can lead to micro- and macrovascular complications. Type 1 diabetes also leads to impaired glucagon production by the pancreatic -cells, whic...

ver descrição completa

Detalhes bibliográficos
Autores: Herrero, P., Oliver, N., Georgiou, P., Bondía Company, Jorge|||0000-0001-7286-3719
Formato: artículo
Fecha de publicación:2017
País:España
Recursos:Universitat Politècnica de València (UPV)
Repositorio:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Idioma:inglés
OAI Identifier:oai:riunet.upv.es:10251/148174
Acesso em linha:https://riunet.upv.es/handle/10251/148174
Access Level:acceso abierto
Palavra-chave:Artificial pancreas
Diabetes
Closed-loop control
Bihormonal control
Glucose control
Insulin delivery
INGENIERIA DE SISTEMAS Y AUTOMATICA
Descrição
Resumo:[EN] Type 1 diabetes is an autoimmune condition characterised by a pancreatic insulin secretion deficit, resulting in high blood glucose concentrations, which can lead to micro- and macrovascular complications. Type 1 diabetes also leads to impaired glucagon production by the pancreatic -cells, which acts as a counter-regulatory hormone to insulin. A closed-loop system for automatic insulin and glucagon delivery, also referred to as an artificial pancreas, has the potential to reduce the self-management burden of type 1 diabetes and reduce the risk of hypo- and hyperglycemia. To date, bihormonal closed-loop systems for glucagon and insulin delivery have been based on two independent controllers. However, in physiology, the secretion of insulin and glucagon in the body is closely interconnected by paracrine and endocrine associations. In this work, we present a novel biologically-inspired glucose control strategy that accounts for such coordination. An in silico study using an FDA-accepted type 1 simulator was performed to evaluate the proposed coordinated control strategy compared to its non-coordinated counterpart, as well as an insulin-only version of the controller. The proposed coordinated strategy achieves a reduction of hyperglycemia without increasing hypoglycemia, when compared to its non-coordinated counterpart.