Effect of applied voltage, duration and repetition frequency of RF pulses for pain relief on temperature spikes and electrical field: A computer modeling study

[EN] Background: The thermal and electrical effects of pulsed radiofrequency (PRF) for pain relief can be controlled by modifying the characteristics of the RF pulses applied. Our goal was to evaluate the influence of such modifications on the thermal and electric performance in tissue. Methods: A c...

ver descrição completa

Detalhes bibliográficos
Autores: Ewertowska, Elzbieta, Mercadal, Borja, Muñoz, Víctor, Ivorra, Antoni, Trujillo Guillen, Macarena|||0000-0003-4145-2188, Berjano, Enrique|||0000-0002-3247-2665
Formato: artículo
Fecha de publicación:2018
País:España
Recursos:Universitat Politècnica de València (UPV)
Repositorio:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Idioma:inglés
OAI Identifier:oai:riunet.upv.es:10251/104542
Acesso em linha:https://riunet.upv.es/handle/10251/104542
Access Level:acceso abierto
Palavra-chave:Computational model
Electroporation
Pain relief
Pulsed radiofrequency
Temperature spike
MATEMATICA APLICADA
TECNOLOGIA ELECTRONICA
Descrição
Resumo:[EN] Background: The thermal and electrical effects of pulsed radiofrequency (PRF) for pain relief can be controlled by modifying the characteristics of the RF pulses applied. Our goal was to evaluate the influence of such modifications on the thermal and electric performance in tissue. Methods: A computational model was developed to compare the temperature and electric field time courses in tissue between a standard clinical protocol (45 V pulses, 20ms duration, 2 Hz repetition frequency) and a new protocol (55 V pulses, 5ms duration, 5 Hz repetition frequency) with a higher applied electric field but a smaller impact on temperature alterations in tissue. The effect of including a temperature controller was assessed. Complementarily, an agar-based experimental model was developed to validate the methodology employed in the computer modelling. Results: The new protocol increased the electric field magnitude reached in the tissue by around +20%, without increasing the temperature. The temperature controller was found to be the fundamental factor in avoiding thermal damage to the tissue and reduced the total number of pulses delivered by around 67%. The experimental results matched moderately well with those obtained from a computer model built especially to mimic the experimental conditions. Conclusions: For the same delivered energy, the new protocol significantly increases the magnitude of the applied electric field, which may be the reason why it is clinically more effective in achieving pain relief.