Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead
Background: In recent years, subcutaneous implantable cardioverter-defibrillator (S-ICD) implants have progressively increased and have been shown to be safe and highly successful, affording low reintervention rates regardless of the technique used. Case summary: We present a case of S-ICD implantat...
| Autores: | , , , |
|---|---|
| Tipo de recurso: | artículo |
| Estado: | Versión publicada |
| Fecha de publicación: | 2019 |
| País: | España |
| Institución: | Universidad de Barcelona |
| Repositorio: | Dipòsit Digital de la UB |
| OAI Identifier: | oai:diposit.ub.edu:2445/163506 |
| Acceso en línea: | https://hdl.handle.net/2445/163506 |
| Access Level: | acceso abierto |
| Palabra clave: | Raigs X Desfibril·ladors cardioversors implantables Fibril·lació ventricular X-rays Implantable cardioverter-defibrillators Ventricular fibrillation |
| id |
ES_ba6716e36b986bb6003e2d7ed9dc4738 |
|---|---|
| oai_identifier_str |
oai:diposit.ub.edu:2445/163506 |
| network_acronym_str |
ES |
| network_name_str |
España |
| repository_id_str |
|
| spelling |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator leadApolo, JoséSan Antonio, RodolfoMont Girbau, LluísTolosana, José M. (José María)Raigs XDesfibril·ladors cardioversors implantablesFibril·lació ventricularX-raysImplantable cardioverter-defibrillatorsVentricular fibrillationBackground: In recent years, subcutaneous implantable cardioverter-defibrillator (S-ICD) implants have progressively increased and have been shown to be safe and highly successful, affording low reintervention rates regardless of the technique used. Case summary: We present a case of S-ICD implantation in a patient diagnosed with idiopathic ventricular fibrillation. In the first follow-up consultation the patient showed appropriate detection parameters in the three configurations. However, chest X-ray revealed lead displacement with a tip migration from the manubrium area of the sternum to the xiphoid process. Discussion: This case highlights the importance of performing at least one chest X-ray during the first weeks after S-ICD implantation, allowing the detection of a problem such as lead displacement, which can lead to undersensing of ventricular arrhythmias or S-ICD oversensing.Oxford University Press2019info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://hdl.handle.net/2445/163506Articles publicats en revistes (Medicina)reponame:Dipòsit Digital de la UBinstname:Universidad de BarcelonaInglésReproducció del document publicat a: https://doi.org/10.1093/ehjcr/ytz189European Heart Journal - Case Reports, 2019, vol. 3, num. 4, p. 1-5https://doi.org/10.1093/ehjcr/ytz189cc-by-nc (c) Apolo, José et al., 2019http://creativecommons.org/licenses/by-nc/3.0/esinfo:eu-repo/semantics/openAccessoai:diposit.ub.edu:2445/1635062026-05-27T06:46:51Z |
| dc.title.none.fl_str_mv |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead |
| title |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead |
| spellingShingle |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead Apolo, José Raigs X Desfibril·ladors cardioversors implantables Fibril·lació ventricular X-rays Implantable cardioverter-defibrillators Ventricular fibrillation |
| title_short |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead |
| title_full |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead |
| title_fullStr |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead |
| title_full_unstemmed |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead |
| title_sort |
Undetected displacement of a subcutaneous implantable cardioverter-defibrillator lead |
| dc.creator.none.fl_str_mv |
Apolo, José San Antonio, Rodolfo Mont Girbau, Lluís Tolosana, José M. (José María) |
| author |
Apolo, José |
| author_facet |
Apolo, José San Antonio, Rodolfo Mont Girbau, Lluís Tolosana, José M. (José María) |
| author_role |
author |
| author2 |
San Antonio, Rodolfo Mont Girbau, Lluís Tolosana, José M. (José María) |
| author2_role |
author author author |
| dc.subject.none.fl_str_mv |
Raigs X Desfibril·ladors cardioversors implantables Fibril·lació ventricular X-rays Implantable cardioverter-defibrillators Ventricular fibrillation |
| topic |
Raigs X Desfibril·ladors cardioversors implantables Fibril·lació ventricular X-rays Implantable cardioverter-defibrillators Ventricular fibrillation |
| description |
Background: In recent years, subcutaneous implantable cardioverter-defibrillator (S-ICD) implants have progressively increased and have been shown to be safe and highly successful, affording low reintervention rates regardless of the technique used. Case summary: We present a case of S-ICD implantation in a patient diagnosed with idiopathic ventricular fibrillation. In the first follow-up consultation the patient showed appropriate detection parameters in the three configurations. However, chest X-ray revealed lead displacement with a tip migration from the manubrium area of the sternum to the xiphoid process. Discussion: This case highlights the importance of performing at least one chest X-ray during the first weeks after S-ICD implantation, allowing the detection of a problem such as lead displacement, which can lead to undersensing of ventricular arrhythmias or S-ICD oversensing. |
| publishDate |
2019 |
| dc.date.none.fl_str_mv |
2019 |
| dc.type.none.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion |
| format |
article |
| status_str |
publishedVersion |
| dc.identifier.none.fl_str_mv |
https://hdl.handle.net/2445/163506 |
| url |
https://hdl.handle.net/2445/163506 |
| dc.language.none.fl_str_mv |
Inglés |
| language_invalid_str_mv |
Inglés |
| dc.relation.none.fl_str_mv |
Reproducció del document publicat a: https://doi.org/10.1093/ehjcr/ytz189 European Heart Journal - Case Reports, 2019, vol. 3, num. 4, p. 1-5 https://doi.org/10.1093/ehjcr/ytz189 |
| dc.rights.none.fl_str_mv |
cc-by-nc (c) Apolo, José et al., 2019 http://creativecommons.org/licenses/by-nc/3.0/es info:eu-repo/semantics/openAccess |
| rights_invalid_str_mv |
cc-by-nc (c) Apolo, José et al., 2019 http://creativecommons.org/licenses/by-nc/3.0/es |
| eu_rights_str_mv |
openAccess |
| dc.format.none.fl_str_mv |
application/pdf |
| dc.publisher.none.fl_str_mv |
Oxford University Press |
| publisher.none.fl_str_mv |
Oxford University Press |
| dc.source.none.fl_str_mv |
Articles publicats en revistes (Medicina) reponame:Dipòsit Digital de la UB instname:Universidad de Barcelona |
| instname_str |
Universidad de Barcelona |
| reponame_str |
Dipòsit Digital de la UB |
| collection |
Dipòsit Digital de la UB |
| repository.name.fl_str_mv |
|
| repository.mail.fl_str_mv |
|
| _version_ |
1869417895167000576 |
| score |
15.300724 |