A Bench Evaluation of Eight Home-Care Ventilators
BACKGROUND: The growing number of patients on home mechanical ventilation has driven considerable progress in the performance and functionality of ventilators, with features comparable with those used in the ICU. However, a publication gap exists in the evaluation and comparison of their performance...
| Autores: | , , , |
|---|---|
| Tipo de recurso: | artículo |
| Estado: | Versión publicada |
| Fecha de publicación: | 2021 |
| País: | España |
| Institución: | INCLIVA |
| Repositorio: | r-INCLIVA. Repositorio Institucional de Producción Científica de INCLIVA |
| OAI Identifier: | oai:incliva.fundanetsuite.com:p16502 |
| Acceso en línea: | https://incliva.portalinvestigacion.com/publicaciones/16502 |
| Access Level: | acceso abierto |
| Palabra clave: | mechanical ventilation noninvasive ventilation breathing mechanics computer simulation home care ventilator performance pressure-time product trigger delay time asynchrony |
| Sumario: | BACKGROUND: The growing number of patients on home mechanical ventilation has driven considerable progress in the performance and functionality of ventilators, with features comparable with those used in the ICU. However, a publication gap exists in the evaluation and comparison of their performance and each ventilator choice depends on machine characteristics defined by manufacturers. METHODS: We bench tested 8 home-care ventilators that are currently available: Monnal T50, EOVE EO-150, Puritan Bennet 560, Weinmann, PrismaVent 50, Trilogy Evo, Astral 150, and Vivo 60 by using an active lung model. These devices were tested under 18 experimental conditions that combined 3 variables: respiratory mechanics, ventilatory mode, and inspiratory muscle effort. The volume delivered, trigger response, pressurization capacity, and synchronization were analyzed. RESULTS: Significant differences were observed in the performance among the devices. Decreased inspiratory muscle effort caused changes in the delivered volume, which worsened the response-to-trigger time, pressurization capacity, and synchronization. Increased pressure support favored the development of asynchronies. All the ventilators developed asynchronies under at least 1 set of conditions, but the EOVE and Trilogy Evo ventilators showed the fewest asynchronies during the experimental conditions studied. CONCLUSIONS: Great variability in terms of technical performance was observed among the 8 home-care ventilators analyzed. Asynchronies became a major issue when home mechanical ventilation was used under higher pressure-support values and lower muscle efforts. Our results may prove to be useful in helping choose the best suited machine based on a patient's clinical therapy needs. |
|---|