Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica

Stent quality control is a critical process. Coronary stents have to be inspected 100% so no defective stent is implanted into a human body. We have developed an optical non-contact measurement instrument that provides stent inspection in an automated and high-speed approach, delivering the user all...

ver descrição completa

Detalhes bibliográficos
Autor: Bermúdez Porras, Carlos
Formato: tesis doctoral
Estado:Versión publicada
Fecha de publicación:2017
País:España
Recursos:CBUC, CESCA
Repositorio:TDR. Tesis Doctorales en Red
OAI Identifier:oai:www.tdx.cat:10803/406098
Acesso em linha:http://hdl.handle.net/10803/406098
https://dx.doi.org/10.5821/dissertation-2117-108237
Access Level:acceso abierto
Palavra-chave:Microscopia
Metrologia òptica
Visió per computador
Inspecció de stents
Inspecció visual
Microscopy
0ptical metrology
Machine vision
Stent inspection
Visual inspection
Àrees temàtiques de la UPC::Ciències de la salut
617
id ES_ab9d4a869ca6be45dfc8b096d0b995c7
oai_identifier_str oai:www.tdx.cat:10803/406098
network_acronym_str ES
network_name_str España
repository_id_str
dc.title.none.fl_str_mv Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica
title Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica
spellingShingle Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica
Bermúdez Porras, Carlos
Microscopia
Metrologia òptica
Visió per computador
Inspecció de stents
Inspecció visual
Microscopy
0ptical metrology
Machine vision
Stent inspection
Visual inspection
Àrees temàtiques de la UPC::Ciències de la salut
617
title_short Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica
title_full Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica
title_fullStr Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica
title_full_unstemmed Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica
title_sort Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òptica
dc.creator.none.fl_str_mv Bermúdez Porras, Carlos
author Bermúdez Porras, Carlos
author_facet Bermúdez Porras, Carlos
author_role author
dc.contributor.none.fl_str_mv Laguarta Bertran, Ferran
Artigas Pursals, Roger
Universitat Politècnica de Catalunya. Departament d'Òptica i Optometria
dc.subject.none.fl_str_mv Microscopia
Metrologia òptica
Visió per computador
Inspecció de stents
Inspecció visual
Microscopy
0ptical metrology
Machine vision
Stent inspection
Visual inspection
Àrees temàtiques de la UPC::Ciències de la salut
617
topic Microscopia
Metrologia òptica
Visió per computador
Inspecció de stents
Inspecció visual
Microscopy
0ptical metrology
Machine vision
Stent inspection
Visual inspection
Àrees temàtiques de la UPC::Ciències de la salut
617
description Stent quality control is a critical process. Coronary stents have to be inspected 100% so no defective stent is implanted into a human body. We have developed an optical non-contact measurement instrument that provides stent inspection in an automated and high-speed approach, delivering the user all the necessary information to assess whether a stent is valid or needs to be rejected. As regards to the instrument design, we have developed and built an optical sensor head that obtains high resolution colour images of the stent surfaces at high speed. Acquired images contain not only contour information of the struts, but also depict the edges roundnesses caused by the surface treatment. This has been achieved by the integration of a triple illumination system in combination with a high precision rotational stage and an area seen camera, which also acquires in line-scan mode. As a result, a patent filed by Sensofar Medical, S.L. protects this invention, which is also being commercialized by the same company. Regarding to the rotational system, different driving configurations have been analysed, and the motion errors that they produce to the images have been studied. Taking this into account, manufacturing tolerances have been established accordingly to guarantee high quality image acquisition. As regards to image acquisition and analysis, we have studied and characterized the difference between the image obtained with a conventional, bright field microscope and the proposed system. Moreover, we have introduced a new method to measure sidewalls critical dimensions, consisting of taking the same, line-scanned images at a certain observation angle. Together with a correction model, 1-micron precision measurements are obtained. To be able to process the measurements in an automated approach, we have developed critical dimension measurement algorithms for the four stent surfaces: abluminals (inner and outer) and luminal (sidewalls). Additionally, we have studied, developed and implemented defect detection algorithms controlled by a sensitivity parameter adjusted by the user. Defects are also classified in a supervised classifier approach by means of a previous training with a defects database. The instrument has also been enabled with tridimensional measurement capabilities, based on coherence scanning interferometry to obtain the topography and roughness of any surface. This technique is also used to characterize defects in 30 in order to reduce the stent rejection rate and thus increasing the process efficiency, as the height of the candidates which maybe are not defects can be also assessed. Besides, this technique can also measure the transparent film thicknesses of drug eluting stents. In this case, due to the cylindrical shape of the samples, measured thicknesses are altered by the local slope at the measuring point, so we have proposed and evaluated two correction models that provide the thickness 30 map corrected automatically. All the capabilities described provide a number of advantages in the stent production and quality control stages. Specifically, objectivity and traceability of these processes are significantly increased, making stents every day more safe devices, reducing the restenosis rate, cardiac accidents or illnesses resulting from an implant malfunction.
publishDate 2017
dc.date.none.fl_str_mv 2017
2017
2017
dc.type.none.fl_str_mv info:eu-repo/semantics/doctoralThesis
info:eu-repo/semantics/publishedVersion
format doctoralThesis
status_str publishedVersion
dc.identifier.none.fl_str_mv http://hdl.handle.net/10803/406098
https://dx.doi.org/10.5821/dissertation-2117-108237
url http://hdl.handle.net/10803/406098
https://dx.doi.org/10.5821/dissertation-2117-108237
dc.language.none.fl_str_mv Catalán
language_invalid_str_mv Catalán
dc.rights.none.fl_str_mv http://creativecommons.org/licenses/by-nc-nd/4.0/
info:eu-repo/semantics/openAccess
rights_invalid_str_mv http://creativecommons.org/licenses/by-nc-nd/4.0/
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 233 p.
application/pdf
application/pdf
dc.publisher.none.fl_str_mv Universitat Politècnica de Catalunya
publisher.none.fl_str_mv Universitat Politècnica de Catalunya
dc.source.none.fl_str_mv TDX (Tesis Doctorals en Xarxa)
reponame:TDR. Tesis Doctorales en Red
instname:CBUC, CESCA
instname_str CBUC, CESCA
reponame_str TDR. Tesis Doctorales en Red
collection TDR. Tesis Doctorales en Red
repository.name.fl_str_mv
repository.mail.fl_str_mv
_version_ 1869416284741959680
spelling Estudi i desenvolupament d'un nou sistema d'inspecció de stents basat en tècniques de metrologia òpticaBermúdez Porras, CarlosMicroscopiaMetrologia òpticaVisió per computadorInspecció de stentsInspecció visualMicroscopy0ptical metrologyMachine visionStent inspectionVisual inspectionÀrees temàtiques de la UPC::Ciències de la salut617Stent quality control is a critical process. Coronary stents have to be inspected 100% so no defective stent is implanted into a human body. We have developed an optical non-contact measurement instrument that provides stent inspection in an automated and high-speed approach, delivering the user all the necessary information to assess whether a stent is valid or needs to be rejected. As regards to the instrument design, we have developed and built an optical sensor head that obtains high resolution colour images of the stent surfaces at high speed. Acquired images contain not only contour information of the struts, but also depict the edges roundnesses caused by the surface treatment. This has been achieved by the integration of a triple illumination system in combination with a high precision rotational stage and an area seen camera, which also acquires in line-scan mode. As a result, a patent filed by Sensofar Medical, S.L. protects this invention, which is also being commercialized by the same company. Regarding to the rotational system, different driving configurations have been analysed, and the motion errors that they produce to the images have been studied. Taking this into account, manufacturing tolerances have been established accordingly to guarantee high quality image acquisition. As regards to image acquisition and analysis, we have studied and characterized the difference between the image obtained with a conventional, bright field microscope and the proposed system. Moreover, we have introduced a new method to measure sidewalls critical dimensions, consisting of taking the same, line-scanned images at a certain observation angle. Together with a correction model, 1-micron precision measurements are obtained. To be able to process the measurements in an automated approach, we have developed critical dimension measurement algorithms for the four stent surfaces: abluminals (inner and outer) and luminal (sidewalls). Additionally, we have studied, developed and implemented defect detection algorithms controlled by a sensitivity parameter adjusted by the user. Defects are also classified in a supervised classifier approach by means of a previous training with a defects database. The instrument has also been enabled with tridimensional measurement capabilities, based on coherence scanning interferometry to obtain the topography and roughness of any surface. This technique is also used to characterize defects in 30 in order to reduce the stent rejection rate and thus increasing the process efficiency, as the height of the candidates which maybe are not defects can be also assessed. Besides, this technique can also measure the transparent film thicknesses of drug eluting stents. In this case, due to the cylindrical shape of the samples, measured thicknesses are altered by the local slope at the measuring point, so we have proposed and evaluated two correction models that provide the thickness 30 map corrected automatically. All the capabilities described provide a number of advantages in the stent production and quality control stages. Specifically, objectivity and traceability of these processes are significantly increased, making stents every day more safe devices, reducing the restenosis rate, cardiac accidents or illnesses resulting from an implant malfunction.El control de qualitat dels stents és un procés crític. El 100% dels stents coronaris han de ser inspeccionats, ja que properament s'han d'implantar en el cos humà. S'ha dissenyat un instrument de mesura sense contacte que permet inspeccionar els stents de forma totalment automatitzada i a alta velocitat, proveint a l'usuari amb tota la informació necessària per jutjar si un stent és vàlid o s'ha de rebutjar. Pel que fa al disseny de l'instrument, s'ha dissenyat i construït un capçal òptic que permet obtenir imatges d'alta resolució i en color de la superfície dels stents a alta velocitat. Les imatges adquirides contenen informació tant del contorn dels struts com de l'arrodoniment de les seves arestes provocat pel procés de tractament superficial. Això s'ha aconseguit incorporant un sistema d'il·luminació triple en combinació amb un sistema de rotació d'alta precisió i una càmera d’àrea que també adquireix en mode line-scan. Com a resultat, s'ha registrat una patent registrada per Sensofar Medical, S.L. que protegeix aquesta invenció i que esta sent actualment explotada per la mateixa empresa. Pel que fa al sistema de rotació, s'han analitzat diferents construccions a nivell d'accionament electromecànic i se n'han estudiat els errors que es produeixen sobre les imatges dels stents. De la mateixa manera, s'han establert unes toleràncies constructives màximes per tal de garantir la captura d'imatge d'alta qualitat. A nivell d'adquisició d'imatge i de l’anàlisi de les mateixes per poder realitzar mesures dimensionals, s'ha estudiat i caracteritzat la diferencia entre les imatges obtingudes amb un microscopi convencional de camp ciar i amb el sistema proposat. A més a més s 'ha plantejat una nova manera de mesurar el gruix de les parets deis stents o sidewalls que consisteix en fer les mateixes imatges line-scan a un cert angle d'observació, que mitjançant uns models de correcció permeten obtenir mesures amb precisió d'una micra. Per poder dur a terme aquest procés de forma automatitzada, s'han desenvolupat els algoritmes de mesura de les dimensions crítiques dels stents a les quatre superfícies dels stents: abluminals (exterior i interior) i luminals (sidewalls). Per altra banda, s'ha estudiat, desenvolupat i implementat la detecció de defectes en base a un paràmetre de sensitivitat ajustat per l'usuari i la seva classificació en un mode supervisat mitjançant un entrenament amb una base de dades de diferents tipus de defectes. A l'instrument s'hi ha integrat la capacitat de mesura tridimensional basada en interferometria de llum blanca per obtenir la topografia i la rugositat de qualsevol superfície o defecte. Aquesta tecnologia també s'utilitza per caracteritzar en tres dimensions els defectes, reduint la taxa de rebuig al poder determinar l'alçada d'aquells candidats que potser no haurien de ser catalogats com a tal, per tant incrementant l'eficiència del procés productiu. Per altra banda, aquesta tècnica també permet mesurar el gruix deis recobriments transparents que es dipositen en els drugeluting stents. En aquest cas , i degut a que aquestes mesures es realitzen sobre una superfície cilíndrica, els gruixos mesurats es troben alterats pel pendent local del punt de mesura, de manera que s'han proposat i avaluat dos models de correcció que obtenen el mapa de gruix tridimensional corregit de forma automàtica. Totes les capacitats descrites constitueixen un conjunt d'avantatges en les etapes de producció i control de qualitat dels stents. Concretament, s'incrementa l'objectivitat i la traçabilitat d'aquests processos, fent que els stents siguin cada vegada més segurs i que per tant hi hagin menys casos de estenosis, accidents cardíacs o malalties derivades d'un mal funcionament de l'implant.DOCTORAT EN ENGINYERIA ÒPTICA (Pla 2007)Universitat Politècnica de CatalunyaLaguarta Bertran, FerranArtigas Pursals, RogerUniversitat Politècnica de Catalunya. Departament d'Òptica i Optometria201720172017info:eu-repo/semantics/doctoralThesisinfo:eu-repo/semantics/publishedVersion233 p.application/pdfapplication/pdfhttp://hdl.handle.net/10803/406098https://dx.doi.org/10.5821/dissertation-2117-108237TDX (Tesis Doctorals en Xarxa)reponame:TDR. Tesis Doctorales en Redinstname:CBUC, CESCACatalánL'accés als continguts d'aquesta tesi queda condicionat a l'acceptació de les condicions d'ús establertes per la següent llicència Creative Commons: http://creativecommons.org/licenses/by-nc-nd/4.0/http://creativecommons.org/licenses/by-nc-nd/4.0/info:eu-repo/semantics/openAccessoai:www.tdx.cat:10803/4060982026-06-14T12:46:07Z
score 15,301603