Description and commissioning of NEXT-MM prototype: first results from operation in a Xenon-Trimethylamine gas mixture
A technical description of NEXT-MM and its commissioning and first performance is reported. Having an active volume of ~ 35 cm drift × 28 cm diameter, it constitutes the largest Micromegas-read TPC operated in Xenon ever constructed, made by a sectorial arrangement of the 4 largest single wafers man...
| Autores: | , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , |
|---|---|
| Tipo de recurso: | artículo |
| Estado: | Versión enviada para evaluación y publicación |
| Fecha de publicación: | 2014 |
| País: | España |
| Institución: | Varias* (Consorci de Biblioteques Universitáries de Catalunya, Centre de Serveis Científics i Acadèmics de Catalunya) |
| Repositorio: | Recercat. Dipósit de la Recerca de Catalunya |
| OAI Identifier: | oai:recercat.cat:10256/16966 |
| Acceso en línea: | http://hdl.handle.net/10256/16966 |
| Access Level: | acceso abierto |
| Palabra clave: | Enginyeria -- Instruments Engineering instruments Detectors de radiació Nuclear counters |
| Sumario: | A technical description of NEXT-MM and its commissioning and first performance is reported. Having an active volume of ~ 35 cm drift × 28 cm diameter, it constitutes the largest Micromegas-read TPC operated in Xenon ever constructed, made by a sectorial arrangement of the 4 largest single wafers manufactured with the Microbulk technique to date. It is equipped with a suitably pixelized readout and with a sufficiently large sensitive volume ( ~ 23 l) so as to contain long ( ~ 20 cm) electron tracks. First results obtained at 1 bar for Xenon and Trymethylamine (Xe-(2%)TMA) mixture are presented. The TPC can accurately reconstruct extended background tracks. An encouraging full-width half-maximum of 11.6 % was obtained for ~ 29 keV gammas without resorting to any data post-processing |
|---|