Calibration of Gamma Ray Impacts in Monolithic-Based Detectors Using Voronoi Diagrams

[EN] Molecular imaging systems, such as positron emission tomography (PET), use detectors providing energy and a 3-D interaction position of a gamma ray within a scintillation block. Monolithic crystals are becoming an alternative to crystal arrays in PET. However, calibration processes are required...

Descripción completa

Detalles Bibliográficos
Autores: González Martínez, Antonio Javier, Freire-López-Fando, Marta, F Sánchez, Benlloch Baviera, Jose María|||0000-0001-6073-1436, Gonzalez-Montoro, Andrea
Tipo de recurso: artículo
Fecha de publicación:2020
País:España
Institución:Universitat Politècnica de València (UPV)
Repositorio:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Idioma:inglés
OAI Identifier:oai:riunet.upv.es:10251/168935
Acceso en línea:https://riunet.upv.es/handle/10251/168935
Access Level:acceso abierto
Palabra clave:Calibration processes
Gamma ray detectors
Monolithic crystals
Positron emission tomography (PET)
SiPM
Voronoi diagrams
Descripción
Sumario:[EN] Molecular imaging systems, such as positron emission tomography (PET), use detectors providing energy and a 3-D interaction position of a gamma ray within a scintillation block. Monolithic crystals are becoming an alternative to crystal arrays in PET. However, calibration processes are required to correct for nonuniformities, mainly produced by the truncation of the scintillation light distribution at the edges. We propose a calibration method based on the Voronoi diagrams. We have used $50 \times 50 \times 15$ mm(3) LYSO blocks coupled to a $12\times 12$ SiPMs array. We have first studied two different interpolation algorithms: 1) weighted average method (WAM) and 2) natural neighbor (NN). We have compared them with an existing calibration based on 1-D monomials. Here, the crystal was laterally black painted and a retroreflector (RR) layer added to the entrance face. The NN exhibited the best results in terms of XY impact position, depth of Interaction, and energy, allowing us to calibrate the whole scintillation volume. Later, the NN interpolation has been tested against different crystal surface treatments, allowing always to correct edge effects. Best energy resolutions were observed when using the reflective layers (12%-14%). However, better linearity was observed with the treatments using black paint. In particular, we obtained the best overall performance when lateral black paint is combined with the RR.