A perfusion chamber for monitoring transepithelial NaCl transport in an in vitro model of the renal tubule

Transepithelial electrical measurements in the renal tubule have provided a better understanding of how kidney regulates electrolyte and water homeostasis through the reabsorption of molecules and ions (e.g., H2O and NaCl). While experiments and measurement techniques using native tissue are difficu...

Descripción completa

Detalles Bibliográficos
Autores: Yeste Lozano, José, Martínez Gimeno, Laura, Illa, Xavi, Laborda, Pablo, Guimerà-Brunet, Anton, Sánchez Marín, Juan P, Villa, Rosa, Giménez, Ignacio
Tipo de recurso: artículo
Estado:Versión aceptada para publicación
Fecha de publicación:2018
País:España
Institución:Consejo Superior de Investigaciones Científicas (CSIC)
Repositorio:DIGITAL.CSIC. Repositorio Institucional del CSIC
OAI Identifier:oai:digital.csic.es:10261/167837
Acceso en línea:http://hdl.handle.net/10261/167837
Access Level:acceso abierto
Palabra clave:Transepithelial electrical resistance
Cell layer capacitance
microfluidic cell culture
transepithelial ion fluxes
sodium reabsorption
renal epithelium
Descripción
Sumario:Transepithelial electrical measurements in the renal tubule have provided a better understanding of how kidney regulates electrolyte and water homeostasis through the reabsorption of molecules and ions (e.g., H2O and NaCl). While experiments and measurement techniques using native tissue are difficult to prepare and to reproduce, cell cultures conducted largely with the Ussing chamber lack the effect of fluid shear stress which is a key physiological stimulus in the renal tubule. To overcome these limitations, we present a modular perfusion chamber for long-term culture of renal epithelial cells under flow that allows the continuous and simultaneous monitoring of both transepithelial electrical parameters and transepithelial NaCl transport. The latter is obtained from electrical conductivity measurements since Na+ and Cl- are the ions that contribute most to the electrical conductivity of a standard physiological solution. The system was validated with epithelial monolayers of raTAL and NRK-52E cells that were characterized electrophysiologically for 5 days under different flow conditions (i.e., apical perfusion, basal, or both). In addition, apical to basal chemical gradients of NaCl (140/70 and 70/140 mM) were imposed in order to demonstrate the feasibility of this methodology for quantifying and monitoring in real time the transepithelial reabsorption of NaCl, which is a primary function of the renal tubule.