Malaria quantitative POC testing using magnetic particles, a paper microfluidic device and a hand-held fluorescence reader
A point-of-care (POC) device is reported for highly sensitive and selective detection of Plasmodium falciparum lactate dehydrogenase (Pf-LDH), a biomarker of malaria infection, based on a single-step magneto-immunoassay, a single-use microfluidic paper device and a customized hand-held fluorescence...
| Autores: | , , , , , , , , |
|---|---|
| Tipo de documento: | artigo |
| Estado: | Versión aceptada para publicación |
| Data de publicação: | 2022 |
| País: | España |
| Recursos: | Universidad de Barcelona |
| Repositório: | Dipòsit Digital de la UB |
| OAI Identifier: | oai:diposit.ub.edu:2445/225668 |
| Acesso em linha: | https://hdl.handle.net/2445/225668 |
| Access Level: | Acceso aberto |
| Palavra-chave: | Plasmodium falciparum Malària Diagnòstic Malaria Diagnosis |
| Resumo: | A point-of-care (POC) device is reported for highly sensitive and selective detection of Plasmodium falciparum lactate dehydrogenase (Pf-LDH), a biomarker of malaria infection, based on a single-step magneto-immunoassay, a single-use microfluidic paper device and a customized hand-held fluorescence reader. The single-step magneto-immunoassay consists in a single 5-min incubation of immuno-modified magnetic particles (c-MAb-MPs), biotinylated detection antibody (bd-MAb), and an enzymatic signal amplifier (Poly-HRP). After on-chip MP concentration and washing, signal generation is achieved by adding a fluorescent enzymatic substrate (QuantaRed). Fluorescence signal is measured using a low-cost customized, portable, and sensible fluorescent detector. The POC affords quantitative Pf-LDH detection in <20 min, with a detection limit of 0.92 ng mL−1 (equivalent to 4.6 parasites μL−1). Furthermore, Pf-LDH quantitation in clinical samples correlates with that provided by the reference ELISA, is more sensitive than a commercial rapid diagnostic test (RDT) and entails little user intervention. These results show that fluorescent paper-based microfluidic devices can be exploited to simplify magneto-immunoassay handling, taking this type of test closer to the requirements of POC testing. |
|---|