Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction
[EN] Radiofrequency catheter ablation (RFCA) is a routine treatment for cardiac arrhythmias. During RFCA, the electrode-tissue interface temperature should be kept below 80°C to avoid thrombus formation. Open-irrigated electrodes facilitate power delivery while keeping low temperatures around the ca...
| Autores: | , , , |
|---|---|
| Tipo de recurso: | artículo |
| Fecha de publicación: | 2016 |
| País: | España |
| Institución: | Universitat Politècnica de València (UPV) |
| Repositorio: | RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia |
| Idioma: | inglés |
| OAI Identifier: | oai:riunet.upv.es:10251/150452 |
| Acceso en línea: | https://riunet.upv.es/handle/10251/150452 |
| Access Level: | acceso abierto |
| Palabra clave: | TECNOLOGIA ELECTRONICA |
| id |
ES_788d1cdb0ebdce23e76cf48797d9bc33 |
|---|---|
| oai_identifier_str |
oai:riunet.upv.es:10251/150452 |
| network_acronym_str |
ES |
| network_name_str |
España |
| repository_id_str |
|
| spelling |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow InteractionGonzález-Suárez, Ana|||0000-0002-1813-4176Berjano, Enrique|||0000-0002-3247-2665Guerra, Jose M.Gerardo-Giorda, LucaTECNOLOGIA ELECTRONICA[EN] Radiofrequency catheter ablation (RFCA) is a routine treatment for cardiac arrhythmias. During RFCA, the electrode-tissue interface temperature should be kept below 80°C to avoid thrombus formation. Open-irrigated electrodes facilitate power delivery while keeping low temperatures around the catheter. No computational model of an open-irrigated electrode in endocardial RFCA accounting for both the saline irrigation flow and the blood motion in the cardiac chamber has been proposed yet. We present the first computational model including both effects at once. The model has been validated against existing experimental results. Computational results showed that the surface lesion width and blood temperature are affected by both the electrode design and the irrigation flow rate. Smaller surface lesion widths and blood temperatures are obtained with higher irrigation flow rate, while the lesion depth is not affected by changing the irrigation flow rate. Larger lesions are obtained with increasing power and the electrode-tissue contact. Also, larger lesions are obtained when electrode is placed horizontally. Overall, the computational findings are in close agreement with previous experimental results providing an excellent tool for future catheter research.This work was supported in part by the Basque Government through the BERC 2014-2017 program and by Spanish Ministry of Economy and Competitiveness MINECO through the BCAM Severo Ochoa excellence accreditation SEV-2013-0323, and also by the Spanish "Plan Estatal de Investigacion, Desarrollo e Innovacion Orientada a los Retos de la Sociedad" under Grant TEC2014-52383-C3-R (TEC2014-52383-C3-1-R) and Grant MTM201569992-R. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. This work was supported in part by the Basque Government through the BERC 2014-2017 program and by Spanish Ministry of Economy and Competitiveness MINECO through the BCAM Severo Ochoa excellence accreditation SEV-2013-0323, and also by the Spanish "Plan Estatal de Investigacion, Desarrollo e Innovacion Orientada a los Retos de la Sociedad" under Grant TEC2014-52383-C3-R (TEC2014-52383-C3-1-R) and Grant MTM2015-69992-RPublic Library of ScienceDepartamento de Ingeniería ElectrónicaEscuela Técnica Superior de Ingeniería Aeroespacial y Diseño IndustrialEscuela Politécnica Superior de GandiaBiosignals & Minimally Invasive Technologies-BioMITGobierno Vasco/Eusko JaurlaritzaMinisterio de Economía y CompetitividadRepositorio Institucional de la Universitat Politècnica de València Riunet20162016-03-03journal articlehttp://purl.org/coar/resource_type/c_6501VoRhttp://purl.org/coar/version/c_970fb48d4fbd8a85info:eu-repo/semantics/articleapplication/pdfhttps://riunet.upv.es/handle/10251/150452reponame:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valénciainstname:Universitat Politècnica de València (UPV)InglésengMinisterio de Economía y Competitividad http://dx.doi.org/10.13039/501100003329 SEV-2013-0323 -Ministerio de Economía y Competitividad http://dx.doi.org/10.13039/501100003329 MTM2015-69992-R MODELIZACION Y APROXIMACION NUMERICA ELECTRO-METABOLICA DEL CEREBROMinisterio de Economía y Competitividad http://dx.doi.org/10.13039/501100003329 TEC2014-52383-C3-1-R TECNOLOGIAS BASADAS EN ENERGIA DE RADIOFRECUENCIA Y MICROONDAS PARA CIRUGIA DE MINIMA INVASIONopen accesshttp://purl.org/coar/access_right/c_abf2Reconocimiento (by)http://creativecommons.org/licenses/by/4.0/info:eu-repo/semantics/openAccessoai:riunet.upv.es:10251/1504522026-06-13T07:49:27Z |
| dc.title.none.fl_str_mv |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction |
| title |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction |
| spellingShingle |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction González-Suárez, Ana|||0000-0002-1813-4176 TECNOLOGIA ELECTRONICA |
| title_short |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction |
| title_full |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction |
| title_fullStr |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction |
| title_full_unstemmed |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction |
| title_sort |
Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction |
| dc.creator.none.fl_str_mv |
González-Suárez, Ana|||0000-0002-1813-4176 Berjano, Enrique|||0000-0002-3247-2665 Guerra, Jose M. Gerardo-Giorda, Luca |
| author |
González-Suárez, Ana|||0000-0002-1813-4176 |
| author_facet |
González-Suárez, Ana|||0000-0002-1813-4176 Berjano, Enrique|||0000-0002-3247-2665 Guerra, Jose M. Gerardo-Giorda, Luca |
| author_role |
author |
| author2 |
Berjano, Enrique|||0000-0002-3247-2665 Guerra, Jose M. Gerardo-Giorda, Luca |
| author2_role |
author author author |
| dc.contributor.none.fl_str_mv |
Departamento de Ingeniería Electrónica Escuela Técnica Superior de Ingeniería Aeroespacial y Diseño Industrial Escuela Politécnica Superior de Gandia Biosignals & Minimally Invasive Technologies-BioMIT Gobierno Vasco/Eusko Jaurlaritza Ministerio de Economía y Competitividad Repositorio Institucional de la Universitat Politècnica de València Riunet |
| dc.subject.none.fl_str_mv |
TECNOLOGIA ELECTRONICA |
| topic |
TECNOLOGIA ELECTRONICA |
| description |
[EN] Radiofrequency catheter ablation (RFCA) is a routine treatment for cardiac arrhythmias. During RFCA, the electrode-tissue interface temperature should be kept below 80°C to avoid thrombus formation. Open-irrigated electrodes facilitate power delivery while keeping low temperatures around the catheter. No computational model of an open-irrigated electrode in endocardial RFCA accounting for both the saline irrigation flow and the blood motion in the cardiac chamber has been proposed yet. We present the first computational model including both effects at once. The model has been validated against existing experimental results. Computational results showed that the surface lesion width and blood temperature are affected by both the electrode design and the irrigation flow rate. Smaller surface lesion widths and blood temperatures are obtained with higher irrigation flow rate, while the lesion depth is not affected by changing the irrigation flow rate. Larger lesions are obtained with increasing power and the electrode-tissue contact. Also, larger lesions are obtained when electrode is placed horizontally. Overall, the computational findings are in close agreement with previous experimental results providing an excellent tool for future catheter research. |
| publishDate |
2016 |
| dc.date.none.fl_str_mv |
2016 2016-03-03 |
| dc.type.none.fl_str_mv |
journal article http://purl.org/coar/resource_type/c_6501 VoR http://purl.org/coar/version/c_970fb48d4fbd8a85 |
| dc.type.openaire.fl_str_mv |
info:eu-repo/semantics/article |
| format |
article |
| dc.identifier.none.fl_str_mv |
https://riunet.upv.es/handle/10251/150452 |
| url |
https://riunet.upv.es/handle/10251/150452 |
| dc.language.none.fl_str_mv |
Inglés eng |
| language_invalid_str_mv |
Inglés |
| language |
eng |
| dc.relation.none.fl_str_mv |
Ministerio de Economía y Competitividad http://dx.doi.org/10.13039/501100003329 SEV-2013-0323 - Ministerio de Economía y Competitividad http://dx.doi.org/10.13039/501100003329 MTM2015-69992-R MODELIZACION Y APROXIMACION NUMERICA ELECTRO-METABOLICA DEL CEREBRO Ministerio de Economía y Competitividad http://dx.doi.org/10.13039/501100003329 TEC2014-52383-C3-1-R TECNOLOGIAS BASADAS EN ENERGIA DE RADIOFRECUENCIA Y MICROONDAS PARA CIRUGIA DE MINIMA INVASION |
| dc.rights.none.fl_str_mv |
open access http://purl.org/coar/access_right/c_abf2 Reconocimiento (by) http://creativecommons.org/licenses/by/4.0/ |
| dc.rights.openaire.fl_str_mv |
info:eu-repo/semantics/openAccess |
| rights_invalid_str_mv |
open access http://purl.org/coar/access_right/c_abf2 Reconocimiento (by) http://creativecommons.org/licenses/by/4.0/ |
| eu_rights_str_mv |
openAccess |
| dc.format.none.fl_str_mv |
application/pdf |
| dc.publisher.none.fl_str_mv |
Public Library of Science |
| publisher.none.fl_str_mv |
Public Library of Science |
| dc.source.none.fl_str_mv |
reponame:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia instname:Universitat Politècnica de València (UPV) |
| instname_str |
Universitat Politècnica de València (UPV) |
| reponame_str |
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia |
| collection |
RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia |
| repository.name.fl_str_mv |
|
| repository.mail.fl_str_mv |
|
| _version_ |
1869411243075305472 |
| score |
15,300719 |