Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement
Monte Carlo efficiency transfer method was used to determine the full energy peak efficiency of a coaxial n-type HPGe detector. The efficiencies calibration curves for three Certificate Reference Materials were determined by efficiency transfer using a 152Eu reference source. The efficiency values o...
| Autores: | , , , , |
|---|---|
| Tipo de recurso: | artículo |
| Fecha de publicación: | 2015 |
| País: | España |
| Institución: | Universidad de Navarra |
| Repositorio: | Dadun. Depósito Académico Digital de la Universidad de Navarra |
| Idioma: | inglés |
| OAI Identifier: | oai:dadun.unav.edu:10171/68219 |
| Acceso en línea: | https://hdl.handle.net/10171/68219 |
| Access Level: | acceso abierto |
| Palabra clave: | HPGe detector Monte Carlo Efficiency transfer Low-background gamma spectrometry |
| id |
ES_71eaa1f16161e4e24602efdb34ca129d |
|---|---|
| oai_identifier_str |
oai:dadun.unav.edu:10171/68219 |
| network_acronym_str |
ES |
| network_name_str |
España |
| repository_id_str |
|
| spelling |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurementMorera-Gómez, Y. (Yasser)|||/items/96c584a2-1fe8-4b6d-82c6-af46ddae2d1cCartas-Aguila, H.A. (Héctor A.)|||/items/7eae5cc2-695b-4a3b-a322-f4d3421354d1Alonso-Hernández, C. M. (Carlos Manuel)|||/items/b3db93e8-f76d-4572-9c1d-df648011bb95Bernal-Castillo, J.L. (José L.)|||/items/cdbdb7f4-b3ea-429c-9d55-b149298f5ae9Guillén-Arruebarrena, A. (Aniel)|||/items/53513f56-614a-42bb-a6d9-72409d873ec7HPGe detectorMonte CarloEfficiency transferLow-background gamma spectrometryMonte Carlo efficiency transfer method was used to determine the full energy peak efficiency of a coaxial n-type HPGe detector. The efficiencies calibration curves for three Certificate Reference Materials were determined by efficiency transfer using a 152Eu reference source. The efficiency values obtained after efficiency transfer were used to calculate the activity concentration of the radionuclides detected in the three materials, which were measured in a low-background gamma spectrometry system. Reported and calculated activity concentration show a good agreement with mean deviations of 5%, which is satisfactory for environmental samples measurement.ElsevierDadun. Depósito Académico Digital Universidad de Navarra20242024-01-0920152015-01-0120152015-01-01journal articlehttp://purl.org/coar/resource_type/c_6501info:eu-repo/semantics/articleapplication/pdfhttps://hdl.handle.net/10171/68219reponame:Dadun. Depósito Académico Digital de la Universidad de Navarrainstname:Universidad de NavarraInglésengopen accesshttp://purl.org/coar/access_right/c_abf2info:eu-repo/semantics/openAccessoai:dadun.unav.edu:10171/682192026-06-21T12:47:57Z |
| dc.title.none.fl_str_mv |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement |
| title |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement |
| spellingShingle |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement Morera-Gómez, Y. (Yasser)|||/items/96c584a2-1fe8-4b6d-82c6-af46ddae2d1c HPGe detector Monte Carlo Efficiency transfer Low-background gamma spectrometry |
| title_short |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement |
| title_full |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement |
| title_fullStr |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement |
| title_full_unstemmed |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement |
| title_sort |
Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement |
| dc.creator.none.fl_str_mv |
Morera-Gómez, Y. (Yasser)|||/items/96c584a2-1fe8-4b6d-82c6-af46ddae2d1c Cartas-Aguila, H.A. (Héctor A.)|||/items/7eae5cc2-695b-4a3b-a322-f4d3421354d1 Alonso-Hernández, C. M. (Carlos Manuel)|||/items/b3db93e8-f76d-4572-9c1d-df648011bb95 Bernal-Castillo, J.L. (José L.)|||/items/cdbdb7f4-b3ea-429c-9d55-b149298f5ae9 Guillén-Arruebarrena, A. (Aniel)|||/items/53513f56-614a-42bb-a6d9-72409d873ec7 |
| author |
Morera-Gómez, Y. (Yasser)|||/items/96c584a2-1fe8-4b6d-82c6-af46ddae2d1c |
| author_facet |
Morera-Gómez, Y. (Yasser)|||/items/96c584a2-1fe8-4b6d-82c6-af46ddae2d1c Cartas-Aguila, H.A. (Héctor A.)|||/items/7eae5cc2-695b-4a3b-a322-f4d3421354d1 Alonso-Hernández, C. M. (Carlos Manuel)|||/items/b3db93e8-f76d-4572-9c1d-df648011bb95 Bernal-Castillo, J.L. (José L.)|||/items/cdbdb7f4-b3ea-429c-9d55-b149298f5ae9 Guillén-Arruebarrena, A. (Aniel)|||/items/53513f56-614a-42bb-a6d9-72409d873ec7 |
| author_role |
author |
| author2 |
Cartas-Aguila, H.A. (Héctor A.)|||/items/7eae5cc2-695b-4a3b-a322-f4d3421354d1 Alonso-Hernández, C. M. (Carlos Manuel)|||/items/b3db93e8-f76d-4572-9c1d-df648011bb95 Bernal-Castillo, J.L. (José L.)|||/items/cdbdb7f4-b3ea-429c-9d55-b149298f5ae9 Guillén-Arruebarrena, A. (Aniel)|||/items/53513f56-614a-42bb-a6d9-72409d873ec7 |
| author2_role |
author author author author |
| dc.contributor.none.fl_str_mv |
Dadun. Depósito Académico Digital Universidad de Navarra |
| dc.subject.none.fl_str_mv |
HPGe detector Monte Carlo Efficiency transfer Low-background gamma spectrometry |
| topic |
HPGe detector Monte Carlo Efficiency transfer Low-background gamma spectrometry |
| description |
Monte Carlo efficiency transfer method was used to determine the full energy peak efficiency of a coaxial n-type HPGe detector. The efficiencies calibration curves for three Certificate Reference Materials were determined by efficiency transfer using a 152Eu reference source. The efficiency values obtained after efficiency transfer were used to calculate the activity concentration of the radionuclides detected in the three materials, which were measured in a low-background gamma spectrometry system. Reported and calculated activity concentration show a good agreement with mean deviations of 5%, which is satisfactory for environmental samples measurement. |
| publishDate |
2015 |
| dc.date.none.fl_str_mv |
2015 2015-01-01 2015 2015-01-01 2024 2024-01-09 |
| dc.type.none.fl_str_mv |
journal article http://purl.org/coar/resource_type/c_6501 |
| dc.type.openaire.fl_str_mv |
info:eu-repo/semantics/article |
| format |
article |
| dc.identifier.none.fl_str_mv |
https://hdl.handle.net/10171/68219 |
| url |
https://hdl.handle.net/10171/68219 |
| dc.language.none.fl_str_mv |
Inglés eng |
| language_invalid_str_mv |
Inglés |
| language |
eng |
| dc.rights.none.fl_str_mv |
open access http://purl.org/coar/access_right/c_abf2 |
| dc.rights.openaire.fl_str_mv |
info:eu-repo/semantics/openAccess |
| rights_invalid_str_mv |
open access http://purl.org/coar/access_right/c_abf2 |
| eu_rights_str_mv |
openAccess |
| dc.format.none.fl_str_mv |
application/pdf |
| dc.publisher.none.fl_str_mv |
Elsevier |
| publisher.none.fl_str_mv |
Elsevier |
| dc.source.none.fl_str_mv |
reponame:Dadun. Depósito Académico Digital de la Universidad de Navarra instname:Universidad de Navarra |
| instname_str |
Universidad de Navarra |
| reponame_str |
Dadun. Depósito Académico Digital de la Universidad de Navarra |
| collection |
Dadun. Depósito Académico Digital de la Universidad de Navarra |
| repository.name.fl_str_mv |
|
| repository.mail.fl_str_mv |
|
| _version_ |
1869410686718705664 |
| score |
15,300719 |