Headspace needle-trap analysis of priority volatile organic compounds from aqueous samples: application to the analysis of natural and waste waters

Combining headspace (HS) sampling with a needle-trap device (NTD) to determine priority volatile organic compounds (VOCs) in water samples results in improved sensitivity and efficiency when compared to conventional static HS sampling. A 22 gauge stainless steel, 51-mm needle packed with Tenax TA an...

Descripción completa

Detalles Bibliográficos
Autores: Alonso Roura, Mònica, Cerdán Andrés, Laura, Godayol Boix, Anna, Anticó i Daró, Ma. Enriqueta, Sánchez Navarro, Juan Manuel
Tipo de recurso: artículo
Estado:Versión aceptada para publicación
Fecha de publicación:2011
País:España
Institución:Varias* (Consorci de Biblioteques Universitáries de Catalunya, Centre de Serveis Científics i Acadèmics de Catalunya)
Repositorio:Recercat. Dipósit de la Recerca de Catalunya
OAI Identifier:oai:recercat.cat:10256/7864
Acceso en línea:http://hdl.handle.net/10256/7864
Access Level:acceso abierto
Palabra clave:Compostos orgànics volàtils
Volatile organic compounds
Aigua -- Anàlisi
Water -- Analysis
Descripción
Sumario:Combining headspace (HS) sampling with a needle-trap device (NTD) to determine priority volatile organic compounds (VOCs) in water samples results in improved sensitivity and efficiency when compared to conventional static HS sampling. A 22 gauge stainless steel, 51-mm needle packed with Tenax TA and Carboxen 1000 particles is used as the NTD. Three different HS-NTD sampling methodologies are evaluated and all give limits of detection for the target VOCs in the ng L−1 range. Active (purge-and-trap) HS-NTD sampling is found to give the best sensitivity but requires exhaustive control of the sampling conditions. The use of the NTD to collect the headspace gas sample results in a combined adsorption/desorption mechanism. The testing of different temperatures for the HS thermostating reveals a greater desorption effect when the sample is allowed to diffuse, whether passively or actively, through the sorbent particles. The limits of detection obtained in the simplest sampling methodology, static HS-NTD (5 mL aqueous sample in 20 mL HS vials, thermostating at 50 °C for 30 min with agitation), are sufficiently low as to permit its application to the analysis of 18 priority VOCs in natural and waste waters. In all cases compounds were detected below regulated levels