Anterior vs. posterior position of dispersive patch during radiofrequency catheter ablation: Insights from in-silico modeling

[EN] Aims: To test the hypothesis that the dispersive patch (DP) location does not significantly affect the current distribution around the catheter tip during radiofrequency catheter ablation (RFCA) but may affect lesions size through differences in impedance due to factors far from the catheter ti...

Descripción completa

Detalles Bibliográficos
Autores: Irastorza, Ramiro M., Maher, Timothy, Barkagan, Michael, Liubasuskas, Rokas, D Avila, Andre, Berjano, Enrique|||0000-0002-3247-2665
Tipo de recurso: artículo
Fecha de publicación:2023
País:España
Institución:Universitat Politècnica de València (UPV)
Repositorio:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Idioma:inglés
OAI Identifier:oai:riunet.upv.es:10251/212372
Acceso en línea:https://riunet.upv.es/handle/10251/212372
Access Level:acceso abierto
Palabra clave:Computer modelling
Dispersive patch
In silico model
Radiofrequency ablation
TECNOLOGIA ELECTRONICA
Descripción
Sumario:[EN] Aims: To test the hypothesis that the dispersive patch (DP) location does not significantly affect the current distribution around the catheter tip during radiofrequency catheter ablation (RFCA) but may affect lesions size through differences in impedance due to factors far from the catheter tip. Methods: An in silico model of RFCA in the posterior left atrium and anterior right ventricle was created using anatomic measurements from patient thoracic computed tomography scans and tested the effect of anterior vs. posterior DP locations on baseline impedance, myocardial power delivery, radiofrequency current path, and predicted lesion size. Results: For posterior left atrium ablation, the baseline impedance, total current delivered, current distribution, and proportion of power delivered to the myocardium were all similar with both anterior and posterior DP locations, resulting in similar RFCA lesion sizes (< 0.2 mm difference). For anterior right ventricular (RV) ablation, an anterior DP location resulted in slightly higher proportion of power delivered to the myocardium and lower baseline impedance leading to slightly larger RFCA lesions (0.6 mm deeper and 0.8 mm wider). Conclusions: An anterior vs. posterior DP location will not meaningfully affect RFCA for posterior left atrial ablation, and the slightly larger lesions predicted with anterior DP location for anterior RV ablation are of unclear clinical significance.