A clinically-oriented computer model for radiofrequency ablation of hepatic tissue with internally cooled wet electrode

[EN] Purpose: To improve the computer modelling of radiofrequency ablation (RFA) by internally cooled wet (ICW) electrodes with added clinically oriented features. Methods: An improved RFA computer model by ICW electrode included: (1) a realistic spatial distribution of the infused saline, and (2) d...

Descripción completa

Detalles Bibliográficos
Autores: Ewertowska, Elzbieta, Quesada-Diez, Rita, Radosevic, A., Andaluz, Anna, Moll, Xavier, García Arnás, F., Burdío Pinilla, Fernando, Berjano, Enrique|||0000-0002-3247-2665, Trujillo Guillen, Macarena|||0000-0003-4145-2188
Tipo de recurso: artículo
Fecha de publicación:2019
País:España
Institución:Universitat Politècnica de València (UPV)
Repositorio:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Idioma:inglés
OAI Identifier:oai:riunet.upv.es:10251/117537
Acceso en línea:https://riunet.upv.es/handle/10251/117537
Access Level:acceso abierto
Palabra clave:Internally cooled wet electrode
Radiofrequency ablation
Computer model
Rehydration
Saline distribution
MATEMATICA APLICADA
TECNOLOGIA ELECTRONICA
Descripción
Sumario:[EN] Purpose: To improve the computer modelling of radiofrequency ablation (RFA) by internally cooled wet (ICW) electrodes with added clinically oriented features. Methods: An improved RFA computer model by ICW electrode included: (1) a realistic spatial distribution of the infused saline, and (2) different domains to distinguish between healthy tissue, saline-infused tumour, and non-infused tumour, under the assumption that infused saline is retained within the tumour boundary. A realistic saline spatial distribution was obtained from an in vivo pig liver study. The computer results were analysed in terms of impedance evolution and coagulation zone (CZ) size, and were compared to the results of clinical trials conducted on 17 patients with the same ICW electrode. Results: The new features added to the model provided computer results that matched well with the clinical results. No roll-offs occurred during the 4-min ablation. CZ transversal diameter (4.10 +/- 0.19 cm) was similar to the computed diameter (4.16 cm). Including the tumour and saline infusion in the model involved (1) a reduction of the initial impedance by 10 - 20 ohm, (2) a delay in roll-off of 20 s and 70 - 100 s, respectively, and (3) 18 - 31% and 22 - 36% larger CZ size, respectively. The saline spatial distribution geometry was also seen to affect roll-off delay and CZ size. Conclusions: Using a three-compartment model and a realistic saline spatial distribution notably improves the match with the outcome of the clinical trials.