Personalised high tibial osteotomy has mechanical safety equivalent to generic device in a case–control in silico clinical trial

Background Despite favourable outcomes relatively few surgeons offer high tibial osteotomy (HTO) as a treatment option for early knee osteoarthritis, mainly due to the difficulty of achieving planned correction and reported soft tissue irritation around the plate used to stablise the osteotomy. To c...

Descripción completa

Detalles Bibliográficos
Autores: MacLeod, Alisdair, Peckham, Nicholas, Serrancolí, Gil|||0000-0001-5034-2445, Rombach, Ines, Hourigan, Patrick, Mandalia, Vipul I., Toms, Andrew, Fregly, Benjamin Jon, Gill, Harinderjit
Tipo de recurso: artículo
Fecha de publicación:2021
País:España
Institución:Universitat Politècnica de Catalunya (UPC)
Repositorio:UPCommons. Portal del coneixement obert de la UPC
Idioma:inglés
OAI Identifier:oai:upcommons.upc.edu:2117/350157
Acceso en línea:https://hdl.handle.net/2117/350157
https://dx.doi.org/10.1038/s43856-021-00001-7
Access Level:acceso abierto
Palabra clave:Biomechanics
Biomecànica
Àrees temàtiques de la UPC::Enginyeria biomèdica::Biomecànica
Descripción
Sumario:Background Despite favourable outcomes relatively few surgeons offer high tibial osteotomy (HTO) as a treatment option for early knee osteoarthritis, mainly due to the difficulty of achieving planned correction and reported soft tissue irritation around the plate used to stablise the osteotomy. To compare the mechanical safety of a new personalised 3D printed high tibial osteotomy (HTO) device, created to overcome these issues, with an existing generic device, a case-control in silico virtual clinical trial was conducted. Methods Twenty-eight knee osteoarthritis patients underwent computed tomography (CT) scanning to create a virtual cohort; the cohort was duplicated to form two arms, Generic and Personalised, on which virtual HTO was performed. Finite element analysis was performed to calculate the stresses in the plates arising from simulated physiological activities at three healing stages. The odds ratio indicative of the relative risk of fatigue failure of the HTO plates between the personalised and generic arms was obtained from a multi-level logistic model. Results Here we show, at 12 weeks post-surgery, the odds ratio indicative of the relative risk of fatigue failure was 0.14 (95%CI 0.01 to 2.73, p¿=¿0.20). Conclusions This novel (to the best of our knowledge) in silico trial, comparing the mechanical safety of a new personalised 3D printed high tibial osteotomy device with an existing generic device, shows that there is no increased risk of failure for the new personalised design compared to the existing generic commonly used device. Personalised high tibial osteotomy can overcome the main technical barriers for this type of surgery, our findings support the case for using this technology for treating early knee osteoarthritis.