Visual acuity and defocus curves with six multifocal intraocular lenses

Purpose The purpose of this prospective, randomized, double-masked, study was to compare the visual performance of patients after bilateral implantation of six different IOLs. Methods The following IOLs were used in the study: SV25T0 (n = 19), ATLISA 809M (n = 18), ATLISA Tri 839MP (n = 19), ZKB00 (...

Descripción completa

Detalles Bibliográficos
Autores: Gil, MA, Varon, C, Cardona, G, Buil, JA
Tipo de recurso: artículo
Estado:Versión publicada
Fecha de publicación:2020
País:España
Institución:Institut d’Investigació Biomèdica Sant Pau (IIB Sant Pau)
Repositorio:r-IIB SANT PAU. Repositorio Institucional de Producción Científica del Instituto de Investigación Biomédica Sant Pau
OAI Identifier:oai:iibsantpau.fundanetsuite.com:p2278
Acceso en línea:https://iibsantpau.fundanetsuite.com/Publicaciones/ProdCientif/PublicacionFrw.aspx?id=2278
http://hdl.handle.net/2117/177660
Access Level:acceso abierto
Palabra clave:Cataract surgery
Defocus curves
Extended depth of focus
Multifocal intraocular lens
Visual acuity
Descripción
Sumario:Purpose The purpose of this prospective, randomized, double-masked, study was to compare the visual performance of patients after bilateral implantation of six different IOLs. Methods The following IOLs were used in the study: SV25T0 (n = 19), ATLISA 809M (n = 18), ATLISA Tri 839MP (n = 19), ZKB00 (n = 20), ZLB00 (n = 20) and Symfony ZXR00 (n = 20). Visual performance was assessed by the monocular distance-corrected visual acuity at 4 m (CDVA), distance-corrected intermediate visual acuity (DCIVA) at 60 cm and distance-corrected near visual acuity (DCNVA) at 40 cm. Binocular defocus curves were obtained for a range of defocus from + 1.50 D to - 4.50 D, in 0.50 D steps. Quality of life was assessed with the VF-14 questionnaire. Results CDVA was better with the Symfony ZXR00 than with the SV25T0 (p = 0.032), ATLISA Tri 839MP (p = 0.032) and ATLISA 809M (p = 0.018). The Symfony ZXR00 offered the best DCIVA, followed by the ZKB00. The best and worst DCNVA results corresponded to the ZLB00, and the SV25T0 and Symfony ZXR00, respectively. Defocus curves at distance were good in all groups, although the Symfony had a wider range of clear vision (- 1.50 D to + 0.50 D), with no decay. For intermediate vision, only the Symfony obtained sharp visual acuity. The ATLISA 809M, ATLISA Tri 839MP and ZLB00 were superior at near distance. Conclusions The extended depth of focus of the Symfony ZXR00 offers a superior range of clear vision at far and intermediate distances than other multifocal designs, with worse results at near distance. Visual outcomes reflect the particular optical, geometrical and power distribution characteristics of each IOL.