Interchangeability of corneal curvature and asphericity measurements provided by three different devices.

AIM: To evaluate the interchangeability of keratometric and asphericity measurements provided by three measurement systems based on different optical principles. METHODS: A total of 40 eyes of 40 patients with a mean age of 34.1y were included. In all cases, a corneal curvature analysis was performe...

Descripción completa

Detalles Bibliográficos
Autores: Pinero, DP, Soto-Negro, R, Rutz-Fortes, P, Perez-Cambrodi, RJ, Fukumitsu, H
Tipo de recurso: artículo
Estado:Versión publicada
Fecha de publicación:2019
País:España
Institución:Fundación para el Fomento de la Investigación Sanitaria y Biomédica de la Comunitat Valenciana (FISABIO)
Repositorio:r-FISABIO. Repositorio Institucional de Producción Científica
OAI Identifier:oai:fisabio.fundanetsuite.com:p7527
Acceso en línea:https://fisabio.portalinvestigacion.com/publicaciones/7527
Access Level:acceso abierto
Palabra clave:astigmatism
corneal asphericity
corneal topography
keratometry
id ES_29de9fa05a49fb3d826ebdfc385e330d
oai_identifier_str oai:fisabio.fundanetsuite.com:p7527
network_acronym_str ES
network_name_str España
repository_id_str
spelling Interchangeability of corneal curvature and asphericity measurements provided by three different devices.Pinero, DPSoto-Negro, RRutz-Fortes, PPerez-Cambrodi, RJFukumitsu, Hastigmatismcorneal asphericitycorneal topographykeratometryAIM: To evaluate the interchangeability of keratometric and asphericity measurements provided by three measurement systems based on different optical principles. METHODS: A total of 40 eyes of 40 patients with a mean age of 34.1y were included. In all cases, a corneal curvature analysis was performed with IOL-Master (IOLM), iDesign 2 (ID2), and Sirius systems (SIR). Differences between instruments for flattest (K1) and steepest (K2) keratometric readings, as well as for magnitude and axis of corneal astigmatism were analyzed. Likewise, differences in asphericity (Q) between SIR and ID2 were also evaluated. RESULTS: Mean differences between devices for K1 were 0.20±0.21 (P<0.001), -0.12±0.36 (P=0.046) and -0.32±0.36 D (P<0.001) for the comparisons IOLM-SIR, IOLM-ID2 and SIR-ID2, respectively. The ranges of agreement for these comparisons between instruments were 0.41, 0.70, and 0.70 D. For K2, mean differences were 0.31±0.33 (P<0.001), -0.08±0.43 (P=0.265) and -0.39±0.38 D (P<0.001), with ranges of agreement of 0.65, 0.84, and 0.74 D. Concerning magnitude of astigmatism, ranges of agreement were in the limit of clinical relevance (0.49 D, P=0.011; 0.55 D, P=0.386; 0.43 D, P=0.05). In contrast, ranges of agreement were clinically relevant for astigmatic axis (26.68°, 33.83° and 18.37°, P=0.121) and for Q between SIR and ID2 (0.16, P<0.001). CONCLUSION: The keratometric corneal power, astigmatic axis and asphericity measurements provide by the three systems evaluated cannot be considered as interchangeable, whereas measurements of corneal astigmatism obtained with SIR and ID2 can be considered as interchangeable for clinical purposes.IJO PRESS2019info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionhttps://fisabio.portalinvestigacion.com/publicaciones/7527International Journal of OphthalmologyISSN: 22223959ISSNe: 22274898reponame:r-FISABIO. Repositorio Institucional de Producción Científicainstname:Fundación para el Fomento de la Investigación Sanitaria y Biomédica de la Comunitat Valenciana (FISABIO)Inglésinfo:eu-repo/semantics/openAccessoai:fisabio.fundanetsuite.com:p75272026-06-11T12:45:17Z
dc.title.none.fl_str_mv Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
title Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
spellingShingle Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
Pinero, DP
astigmatism
corneal asphericity
corneal topography
keratometry
title_short Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
title_full Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
title_fullStr Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
title_full_unstemmed Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
title_sort Interchangeability of corneal curvature and asphericity measurements provided by three different devices.
dc.creator.none.fl_str_mv Pinero, DP
Soto-Negro, R
Rutz-Fortes, P
Perez-Cambrodi, RJ
Fukumitsu, H
author Pinero, DP
author_facet Pinero, DP
Soto-Negro, R
Rutz-Fortes, P
Perez-Cambrodi, RJ
Fukumitsu, H
author_role author
author2 Soto-Negro, R
Rutz-Fortes, P
Perez-Cambrodi, RJ
Fukumitsu, H
author2_role author
author
author
author
dc.subject.none.fl_str_mv astigmatism
corneal asphericity
corneal topography
keratometry
topic astigmatism
corneal asphericity
corneal topography
keratometry
description AIM: To evaluate the interchangeability of keratometric and asphericity measurements provided by three measurement systems based on different optical principles. METHODS: A total of 40 eyes of 40 patients with a mean age of 34.1y were included. In all cases, a corneal curvature analysis was performed with IOL-Master (IOLM), iDesign 2 (ID2), and Sirius systems (SIR). Differences between instruments for flattest (K1) and steepest (K2) keratometric readings, as well as for magnitude and axis of corneal astigmatism were analyzed. Likewise, differences in asphericity (Q) between SIR and ID2 were also evaluated. RESULTS: Mean differences between devices for K1 were 0.20±0.21 (P<0.001), -0.12±0.36 (P=0.046) and -0.32±0.36 D (P<0.001) for the comparisons IOLM-SIR, IOLM-ID2 and SIR-ID2, respectively. The ranges of agreement for these comparisons between instruments were 0.41, 0.70, and 0.70 D. For K2, mean differences were 0.31±0.33 (P<0.001), -0.08±0.43 (P=0.265) and -0.39±0.38 D (P<0.001), with ranges of agreement of 0.65, 0.84, and 0.74 D. Concerning magnitude of astigmatism, ranges of agreement were in the limit of clinical relevance (0.49 D, P=0.011; 0.55 D, P=0.386; 0.43 D, P=0.05). In contrast, ranges of agreement were clinically relevant for astigmatic axis (26.68°, 33.83° and 18.37°, P=0.121) and for Q between SIR and ID2 (0.16, P<0.001). CONCLUSION: The keratometric corneal power, astigmatic axis and asphericity measurements provide by the three systems evaluated cannot be considered as interchangeable, whereas measurements of corneal astigmatism obtained with SIR and ID2 can be considered as interchangeable for clinical purposes.
publishDate 2019
dc.date.none.fl_str_mv 2019
dc.type.none.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.none.fl_str_mv https://fisabio.portalinvestigacion.com/publicaciones/7527
url https://fisabio.portalinvestigacion.com/publicaciones/7527
dc.language.none.fl_str_mv Inglés
language_invalid_str_mv Inglés
dc.rights.none.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.publisher.none.fl_str_mv IJO PRESS
publisher.none.fl_str_mv IJO PRESS
dc.source.none.fl_str_mv International Journal of Ophthalmology
ISSN: 22223959
ISSNe: 22274898
reponame:r-FISABIO. Repositorio Institucional de Producción Científica
instname:Fundación para el Fomento de la Investigación Sanitaria y Biomédica de la Comunitat Valenciana (FISABIO)
instname_str Fundación para el Fomento de la Investigación Sanitaria y Biomédica de la Comunitat Valenciana (FISABIO)
reponame_str r-FISABIO. Repositorio Institucional de Producción Científica
collection r-FISABIO. Repositorio Institucional de Producción Científica
repository.name.fl_str_mv
repository.mail.fl_str_mv
_version_ 1869405035142578176
score 15,812429