Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study

[EN] Background: Preclinical studies on radiofrequency (RF) cardiac ablation (RFCA) use very small temperature sensors in specific positions in the tissue subjected to RF heating. Despite the sensors' small size, the proximity to the ablation electrode and the extremely high thermal gradien...

Descripción completa

Detalles Bibliográficos
Autores: Luis, Berjano, Enrique|||0000-0002-3247-2665, González-Suárez, Ana|||0000-0002-1813-4176, Pérez, Juan J|||0000-0001-8486-8699
Tipo de recurso: artículo
Fecha de publicación:2023
País:España
Institución:Universitat Politècnica de València (UPV)
Repositorio:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
Idioma:inglés
OAI Identifier:oai:riunet.upv.es:10251/193321
Acceso en línea:https://riunet.upv.es/handle/10251/193321
Access Level:acceso abierto
Palabra clave:Cardiac ablation
Computer modeling
Pre-clinical model
Radiofrequency ablation
Temperature sensors
TECNOLOGIA ELECTRONICA
id ES_27da37f5d936fcddf1007293deaed7e4
oai_identifier_str oai:riunet.upv.es:10251/193321
network_acronym_str ES
network_name_str España
repository_id_str
spelling Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling StudyLuisBerjano, Enrique|||0000-0002-3247-2665González-Suárez, Ana|||0000-0002-1813-4176Pérez, Juan J|||0000-0001-8486-8699Cardiac ablationComputer modelingPre-clinical modelRadiofrequency ablationTemperature sensorsTECNOLOGIA ELECTRONICA[EN] Background: Preclinical studies on radiofrequency (RF) cardiac ablation (RFCA) use very small temperature sensors in specific positions in the tissue subjected to RF heating. Despite the sensors' small size, the proximity to the ablation electrode and the extremely high thermal gradient around the electrode means that the presence of the temperature sensors could distort the temperatures recorded. Our objective was to assess the thermal impact of intra-tissue temperature sensors during RFCA. Methods: 3D RFCA models were built including different temperature sensors based on fiber optics and T-type thermocouples. Constant power ablation was simulated for 10 s. Results: The results showed that the disturbance caused by the presence of the T-type thermocouples was considerably greater (one order of magnitude) than that caused by the optical fibers. The closer the sensor was to the ablation electrode, the greater the greater the disturbance was and the more it increased with time in sensors more than 3 mm deep. The fiber optic measurements always slightly underestimated (<0.2 'C) the tissue temperature that would exist without the sensors, while the disturbance caused by the T-type thermocouples did not always result in underestimation but depended on the depth of the sensors parallel to the catheter. Conclusions: The presence of thermocouples inserted into the tissue close to the RF ablation electrode involves a disturbance that could affect the measured temperature value, although it does not substantially alter the shape and size of the thermal lesion. Optical fibers cause much less disturbance, possibly due to the absence of internal metal parts that favor heat conduction.This research was funded by the Spanish Ministerio de Ciencia e Innovación/Agencia Estatal de Investigación (MCIN/AEI/10.13039/501100011033) by Grant number PID2022-136273OBC31 and PID2022-136273OA-C33.MDPI AGDepartamento de Ingeniería ElectrónicaEscuela Técnica Superior de Ingeniería Aeroespacial y Diseño IndustrialEscuela Técnica Superior de Ingeniería IndustrialEscuela Politécnica Superior de GandiaBiosignals & Minimally Invasive Technologies-BioMITAgencia Estatal de InvestigaciónRepositorio Institucional de la Universitat Politècnica de València Riunet20232023-04-11journal articlehttp://purl.org/coar/resource_type/c_6501VoRhttp://purl.org/coar/version/c_970fb48d4fbd8a85info:eu-repo/semantics/articleapplication/pdfhttps://riunet.upv.es/handle/10251/193321reponame:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valénciainstname:Universitat Politècnica de València (UPV)InglésengAgencia Estatal de Investigación http://dx.doi.org/10.13039/501100011033 PID2022-136273OBC31Agencia Estatal de Investigación http://dx.doi.org/10.13039/501100011033 PID2022-136273OA-C33open accesshttp://purl.org/coar/access_right/c_abf2Reconocimiento (by)http://creativecommons.org/licenses/by/4.0/info:eu-repo/semantics/openAccessoai:riunet.upv.es:10251/1933212026-06-13T07:49:27Z
dc.title.none.fl_str_mv Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
title Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
spellingShingle Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
Luis
Cardiac ablation
Computer modeling
Pre-clinical model
Radiofrequency ablation
Temperature sensors
TECNOLOGIA ELECTRONICA
title_short Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
title_full Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
title_fullStr Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
title_full_unstemmed Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
title_sort Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
dc.creator.none.fl_str_mv Luis
Berjano, Enrique|||0000-0002-3247-2665
González-Suárez, Ana|||0000-0002-1813-4176
Pérez, Juan J|||0000-0001-8486-8699
author Luis
author_facet Luis
Berjano, Enrique|||0000-0002-3247-2665
González-Suárez, Ana|||0000-0002-1813-4176
Pérez, Juan J|||0000-0001-8486-8699
author_role author
author2 Berjano, Enrique|||0000-0002-3247-2665
González-Suárez, Ana|||0000-0002-1813-4176
Pérez, Juan J|||0000-0001-8486-8699
author2_role author
author
author
dc.contributor.none.fl_str_mv Departamento de Ingeniería Electrónica
Escuela Técnica Superior de Ingeniería Aeroespacial y Diseño Industrial
Escuela Técnica Superior de Ingeniería Industrial
Escuela Politécnica Superior de Gandia
Biosignals & Minimally Invasive Technologies-BioMIT
Agencia Estatal de Investigación
Repositorio Institucional de la Universitat Politècnica de València Riunet
dc.subject.none.fl_str_mv Cardiac ablation
Computer modeling
Pre-clinical model
Radiofrequency ablation
Temperature sensors
TECNOLOGIA ELECTRONICA
topic Cardiac ablation
Computer modeling
Pre-clinical model
Radiofrequency ablation
Temperature sensors
TECNOLOGIA ELECTRONICA
description [EN] Background: Preclinical studies on radiofrequency (RF) cardiac ablation (RFCA) use very small temperature sensors in specific positions in the tissue subjected to RF heating. Despite the sensors' small size, the proximity to the ablation electrode and the extremely high thermal gradient around the electrode means that the presence of the temperature sensors could distort the temperatures recorded. Our objective was to assess the thermal impact of intra-tissue temperature sensors during RFCA. Methods: 3D RFCA models were built including different temperature sensors based on fiber optics and T-type thermocouples. Constant power ablation was simulated for 10 s. Results: The results showed that the disturbance caused by the presence of the T-type thermocouples was considerably greater (one order of magnitude) than that caused by the optical fibers. The closer the sensor was to the ablation electrode, the greater the greater the disturbance was and the more it increased with time in sensors more than 3 mm deep. The fiber optic measurements always slightly underestimated (<0.2 'C) the tissue temperature that would exist without the sensors, while the disturbance caused by the T-type thermocouples did not always result in underestimation but depended on the depth of the sensors parallel to the catheter. Conclusions: The presence of thermocouples inserted into the tissue close to the RF ablation electrode involves a disturbance that could affect the measured temperature value, although it does not substantially alter the shape and size of the thermal lesion. Optical fibers cause much less disturbance, possibly due to the absence of internal metal parts that favor heat conduction.
publishDate 2023
dc.date.none.fl_str_mv 2023
2023-04-11
dc.type.none.fl_str_mv journal article
http://purl.org/coar/resource_type/c_6501
VoR
http://purl.org/coar/version/c_970fb48d4fbd8a85
dc.type.openaire.fl_str_mv info:eu-repo/semantics/article
format article
dc.identifier.none.fl_str_mv https://riunet.upv.es/handle/10251/193321
url https://riunet.upv.es/handle/10251/193321
dc.language.none.fl_str_mv Inglés
eng
language_invalid_str_mv Inglés
language eng
dc.relation.none.fl_str_mv Agencia Estatal de Investigación http://dx.doi.org/10.13039/501100011033 PID2022-136273OBC31
Agencia Estatal de Investigación http://dx.doi.org/10.13039/501100011033 PID2022-136273OA-C33
dc.rights.none.fl_str_mv open access
http://purl.org/coar/access_right/c_abf2
Reconocimiento (by)
http://creativecommons.org/licenses/by/4.0/
dc.rights.openaire.fl_str_mv info:eu-repo/semantics/openAccess
rights_invalid_str_mv open access
http://purl.org/coar/access_right/c_abf2
Reconocimiento (by)
http://creativecommons.org/licenses/by/4.0/
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv MDPI AG
publisher.none.fl_str_mv MDPI AG
dc.source.none.fl_str_mv reponame:RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
instname:Universitat Politècnica de València (UPV)
instname_str Universitat Politècnica de València (UPV)
reponame_str RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
collection RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia
repository.name.fl_str_mv
repository.mail.fl_str_mv
_version_ 1869404911649685504