Low dose response results and detection limits of the EURADOS intercomparison IC2021area for passive H*(10) area dosemeters
The EURADOS intercomparison IC2021area with its main focus on low dose measurements was carried out between May 2021 and April 2022 for passive H*(10) area dosimetry systems. Three measurement conditions were provided at locations of the Karlsruhe Institute of Technology: 3-months indoor, 3-months o...
| Autores: | , , , , , |
|---|---|
| Formato: | artículo |
| Fecha de publicación: | 2023 |
| País: | España |
| Recursos: | Universitat Politècnica de Catalunya (UPC) |
| Repositorio: | UPCommons. Portal del coneixement obert de la UPC |
| Idioma: | inglés |
| OAI Identifier: | oai:upcommons.upc.edu:2117/388373 |
| Acesso em linha: | https://hdl.handle.net/2117/388373 https://dx.doi.org/10.1016/j.radmeas.2023.106935 |
| Access Level: | acceso abierto |
| Palavra-chave: | Dosimeters EURADOS Intercomparison Area dosimetry Passive dosemeter Dosemeter response Background dose Detection limit Dosímetres Àrees temàtiques de la UPC::Energies |
| Resumo: | The EURADOS intercomparison IC2021area with its main focus on low dose measurements was carried out between May 2021 and April 2022 for passive H*(10) area dosimetry systems. Three measurement conditions were provided at locations of the Karlsruhe Institute of Technology: 3-months indoor, 3-months outdoor and 6-months outdoor. The challenge for the 65 participating dosimetry systems of this intercomparison was measuring additionally irradiated low dose 137Cs gamma radiation of 150 µSv and 300 µSv. Despite the comparably high background dose values between about 200 µSv and 1600 µSv, more than 90% of the resulting response values of the irradiated dosemeters were within the recommended ISO 14146 response limits. The calculated detection limits of the participating dosimetry systems were in most cases clearly below the desired value of 150 µSv of this intercomparison. |
|---|