Recent developments of the 1 MV AMS facility at the Centro Nacional de Aceleradores

The Centro Nacional de Aceleradores (CNA) hosts a 1 MV accelerator mass spectrometry (AMS) apparatus since September 2005. In order to improve its overall performance, several updates have been made on the existing facility during the last 10 years of operation. In this paper, two modifications cond...

ver descrição completa

Detalhes bibliográficos
Autores: Scognamiglio, Grazia, Chamizo Calvo, Elena, López Gutiérrez, José María, Müller, A.M., Padilla, S., Santos, F. Javier, Vivo Vilches, Carlos, García León, Manuel
Formato: artículo
Estado:Versión publicada
Fecha de publicación:2016
País:España
Recursos:Universidad de Sevilla (US)
Repositorio:idUS. Depósito de Investigación de la Universidad de Sevilla
OAI Identifier:oai:idus.us.es:11441/148034
Acesso em linha:https://hdl.handle.net/11441/148034
https://doi.org/10.1016/j.nimb.2016.03.033
Access Level:acceso abierto
Palavra-chave:AMS
Low-energy AMS
CNA
Descrição
Resumo:The Centro Nacional de Aceleradores (CNA) hosts a 1 MV accelerator mass spectrometry (AMS) apparatus since September 2005. In order to improve its overall performance, several updates have been made on the existing facility during the last 10 years of operation. In this paper, two modifications conducted in 2015 will be described. To increase the transmission of the ions through the accelerator, the stripping gas on the 1 MV CNA machine was changed from Ar to He. The measured maximum transmission for almost every isotope results to be higher, especially for heavy masses: for instance, in the case of uranium in the 3+ charge state, the transmission increased from 11% with Ar gas to about 38% with He gas. The second advance consisted of the substitution of the existing gas ionization chamber with a new one provided by ETH Zurich. The ETH detector features with its miniaturized design and is optimized for low energy AMS (i.e. very low electronic noise and efficient charge collection). As the electronic noise is the most important contribution to the resolution for light ions, the total energy resolution has been reduced by 15% in the case of 10Be, allowing a better discrimination against its isobar, 10B. For the heaviest radionuclides where the quality of the spectra is determined by the charge carrier production in the gas, the resolution for 2.7 MeV uranium ions was improved by 30%, probably due to a more efficient charge collection.