A simple method for efficiency calibration of HPGe detectors in gamma-spectrometric measurements
In this paper a simple, rapid and general method for g-ray efficiency calibration of Ge detectors for environmental samples is presented. This method is based on the use of an active natural solid sample with several g-emissions (in our case, 226Ra) as the calibrating matrix for determining the full...
| Autores: | , , , , , , |
|---|---|
| Tipo de recurso: | artículo |
| Estado: | Versión publicada |
| Fecha de publicación: | 2001 |
| País: | España |
| Institución: | Universidad de Sevilla (US) |
| Repositorio: | idUS. Depósito de Investigación de la Universidad de Sevilla |
| OAI Identifier: | oai:idus.us.es:11441/146476 |
| Acceso en línea: | https://hdl.handle.net/11441/146476 https://doi.org/10.1016/S0969-806X(01)00290-0 |
| Access Level: | acceso abierto |
| Palabra clave: | Gamma-spectrometry Efficiency-calibration Self-absorption Sample-height |
| Sumario: | In this paper a simple, rapid and general method for g-ray efficiency calibration of Ge detectors for environmental samples is presented. This method is based on the use of an active natural solid sample with several g-emissions (in our case, 226Ra) as the calibrating matrix for determining the full energy peak efficiency (FEPE) ec vs g-emission energy Eg and the sample height h in a counting cylindrical geometry. The 226Ra activity concentration is determined by a-particle spectrometry, a method that has previously been validated |
|---|