Safety and Efficacy of TAVR With a Pressure Sensor and Pacing Guidewire
The SavvyWire (OpSens Inc) is a 0.035-inch preshaped guidewire with dedicated pacing properties and a distal pressure sensor allowing for continuous hemodynamic pressure monitoring. This study sought to determine the efficacy and safety of the guidewire during transcatheter aortic valve replacement...
| Autores: | , , , , , , , , , , , , |
|---|---|
| Tipo de recurso: | artículo |
| Fecha de publicación: | 2023 |
| País: | España |
| Institución: | Universitat Autònoma de Barcelona |
| Repositorio: | Dipòsit Digital de Documents de la UAB |
| Idioma: | inglés |
| OAI Identifier: | oai:ddd.uab.cat:296634 |
| Acceso en línea: | https://ddd.uab.cat/record/296634 https://dx.doi.org/urn:doi:10.1016/j.jcin.2023.10.035 |
| Access Level: | acceso abierto |
| Palabra clave: | Aortic stenosis Hemodynamic assessment Left ventricular pacing Transcatheter aortic valve replacement |
| Sumario: | The SavvyWire (OpSens Inc) is a 0.035-inch preshaped guidewire with dedicated pacing properties and a distal pressure sensor allowing for continuous hemodynamic pressure monitoring. This study sought to determine the efficacy and safety of the guidewire during transcatheter aortic valve replacement (TAVR) procedures. This prospective, multicenter clinical study included patients with severe aortic stenosis undergoing TAVR in 8 European centers. The primary efficacy endpoint was defined as effective left ventricular rapid pacing runs with the guidewire translating into a significant systemic pressure drop (below 60 mm Hg). The safety outcome included the absence of major procedural complications related to the guidewire. A total of 121 patients (mean age: 82.2 ± 5.9 years, 50% women) were included in the study, and 119 (98.3%) patients were finally treated with the study device. A balloon-expandable valve was implanted in 45 (37.8%) patients. Predilatation and postdilatation were performed in 89 (74.8%) and 14 (11.8%) patients, respectively. The primary efficacy endpoint was achieved in 116 (98.3%) patients, and the mean aortic systolic arterial pressure achieved during rapid pacing was 46.6 ± 11.3 mm Hg. Hemodynamic assessment with the use of the OptoMonitor 3 (OpSens Inc) without additional catheter exchange was achieved in 117 (99.2%) patients. The safety endpoint was achieved in 117 (99.2%) patients. No procedural mortality, stroke, or ventricular perforation was reported. The use of the guidewire during TAVR procedures appeared to be efficacious and safe. This device could help minimize interventions during the procedure and improve the clinical decision making after transcatheter heart valve deployment. (SavvyWire Efficacy and Safety in Transcatheter Aortic Valve Implantation Procedures [SAFE-TAVI]; NCT05492383). |
|---|