Manganese Phthalocyanine as a Biomimetic Electrocatalyst for Phenols in the Development of an Amperometric Sensor
The construction of an amperometric sensor for phenolic compounds using a carbon paste electrode modified with manganese phthalocyanine and histidine is reported. Under optimized conditions, at 0 mV vs Ag/AgCl in 0.1 mol dm(-3) phosphate buffer solutions (pH 7.0) containing 250 mu mol dm(-3) H2O2, a...
| Autores: | , , , , , , |
|---|---|
| Tipo de recurso: | artículo |
| Estado: | Versión publicada |
| Fecha de publicación: | 2009 |
| País: | Brasil |
| Institución: | Universidade Estadual Paulista (UNESP) |
| Repositorio: | Repositório Institucional da UNESP |
| Idioma: | inglés |
| OAI Identifier: | oai:repositorio.unesp.br:11449/40268 |
| Acceso en línea: | http://dx.doi.org/10.1590/S0103-50532009000600025 http://hdl.handle.net/11449/40268 |
| Access Level: | acceso abierto |
| Palabra clave: | manganese phthalocyanine phenolic compound amperometric sensor |
| Sumario: | The construction of an amperometric sensor for phenolic compounds using a carbon paste electrode modified with manganese phthalocyanine and histidine is reported. Under optimized conditions, at 0 mV vs Ag/AgCl in 0.1 mol dm(-3) phosphate buffer solutions (pH 7.0) containing 250 mu mol dm(-3) H2O2, a linear response for catechol from 20 up to 130 mu mol dm(-3) was obtained with a sensitivity of 0.56 (+/- 0.02) mu A dm(3) mu mol(-1) cm(-2). The detection limit was 1.1 mu mol dm(-3) and the response time was 2 s. The sensor presented stable response during 40 successive determinations. The repeatability, evaluated in terms of relative standard deviation, was 3.6% for n = 5 and [catechol] = 10 mu mol dm(-3). The sensor responses for other phenolic compounds were also investigated in details. Finally, the applicability of the sensor was tested in a pharmaceutical containing dopamine and the results compared favorably with those obtained with the official method. |
|---|