In Vivo accuracy of two electronic foramen locators based on different operation systems

The aim of this study was to compare in vivo the accuracy of two electronic foramen locators (EFLs) based on different operation systems - Root ZX and Propex II. Ten healthy adult patients needing premolar extractions due to orthodontic reasons participated in the study, providing a sample of 17 non...

Descripción completa

Detalles Bibliográficos
Autores: Vasconcelos, Bruno Carvalho de, Araújo, Rebeca Bastos Rocha, Silva, Francisco Cláudio Fernandes Alves e, Luna-Cruz, Suyane Maria, Duarte, Marco Antonio Hungaro, Fernandes, Carlos Augusto de Oliveira
Tipo de recurso: artículo
Estado:Versión publicada
Fecha de publicación:2014
País:Brasil
Institución:Universidade Federal do Ceará (UFC)
Repositorio:Repositório Institucional da Universidade Federal do Ceará (UFC)
Idioma:inglés
OAI Identifier:oai:repositorio.ufc.br:riufc/10986
Acceso en línea:http://www.repositorio.ufc.br/handle/riufc/10986
Access Level:acceso abierto
Palabra clave:Cavidade Pulpar
Ápice Dentário
Descripción
Sumario:The aim of this study was to compare in vivo the accuracy of two electronic foramen locators (EFLs) based on different operation systems - Root ZX and Propex II. Ten healthy adult patients needing premolar extractions due to orthodontic reasons participated in the study, providing a sample of 17 noncarious, non-restored, vital teeth (n= 24 canals). After coronal access preparation and cervical preflaring and prior to tooth extraction, the root canal length was measured alternating the two EFLs. All measurements were performed with K-files well fitted to the canal diameter at the level that each EFL indicated the apical foramen in their display (APEX or 0.0). The last K-file were fixed in place with cyanoacrylate, the tooth was extracted, and the apical 4 mm of each root were resected to measure the distance between the file tip and the apical foramen. The mean errors based on the absolute values of discrepancies were 0.30 ± 0.29 mm (Root ZX) and 0.32 ± 0.27 mm (Propex II). Analysis by the Wilcoxon test for paired samples showed no statistically significant differences between the electronic canal measurements performed with the EFLs (p=0.587). The apical foramen was accurately located in 75% (Root ZX) and 66.7% (Propex II) of the cases, considering a ±0.5 mm error margin, with no statistically significant difference by the chi-square test. Despite having different measurement mechanisms, both EFLs were capable of locating the apical foramen with high accuracy in vivo. Under the tested clinical conditions, Root ZX and Propex II displayed similar results.