Stress analysis of cantilever-fixed partial denture connector design using the finite element method

Purpose: To analyze the stress distribution on a cantilever-fixed partial denture after simulation of maximum mastication loads in order to optimize its design. Methods: A cantilever-fixed partial denture framework was designed in the CAD-CAM system Everest®Kavo v2.0 using two materials, titanium an...

ver descrição completa

Detalhes bibliográficos
Autores: Correia, André Ricardo Maia, Sampaio Fernandes, João Carlos, Reis Campos, José Carlos, Vaz, Mário Augusto Pires, Ramos, Nuno Viriato Marques
Tipo de documento: artigo
Estado:Versão publicada
Data de publicação:2009
País:Brasil
Recursos:Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS)
Repositório:Revista odonto ciência (Online)
Idioma:inglês
OAI Identifier:oai:ojs.revistaseletronicas.pucrs.br:article/5115
Acesso em linha:https://revistaseletronicas.pucrs.br/fo/article/view/5115
Access Level:Acceso aberto
Palavra-chave:Fixed partial denture
Yttria-stabilized tetragonal zirconia polycrystals
ceramic
titanium
dental stress analysis
Biomechanics
Descrição
Resumo:Purpose: To analyze the stress distribution on a cantilever-fixed partial denture after simulation of maximum mastication loads in order to optimize its design. Methods: A cantilever-fixed partial denture framework was designed in the CAD-CAM system Everest®Kavo v2.0 using two materials, titanium and zirconium, with connectors of 5.28 mm2 and 9.05 mm2, respectively. A finite element model was built for stress analysis using simulations of mastication load. Results: For zirconia, only the molar cantilever with the smaller connector area and a 0.5-mm fillet exceeded the considered threshold resistance value of 575 MPa. All the other designs yielded resistances below this value. For titanium, only cantilevers with 9.05 mm2 connector area and fillets of 1 or 1.4 mm presented stress values inferior to titanium yield strengh. Conclusion: Within the limitations of this study, it can be concluded that titanium cantilever-fixed partial denture frameworks with a 5.28 mm2 connector area cannot support maximum mastication loads; frameworks of this material require larger connectors with fillets introduced in the gingival embrasure. Zirconia, however, supports maximum bite forces in most situations with both molar and premolar design cantilevers. Precaution should be taken when dealing with smaller connectors of 5.28 mm2.