Validation of a gas chromatography-mass spectrometry method for the analysis of sterol oxidation products in serum
A validated gas chromatography-mass spectrometry (GC-MS) detection method for the quantitative analysis of sterol oxidation products (SOPs) in serum is described. After a lipid extraction procedure with chloroform-methanol, a cold saponification and purification by solid phase extraction, oxysterols...
| Autores: | , , , |
|---|---|
| Tipo de recurso: | artículo |
| Fecha de publicación: | 2008 |
| País: | España |
| Institución: | Universidad de Navarra |
| Repositorio: | Dadun. Depósito Académico Digital de la Universidad de Navarra |
| Idioma: | inglés |
| OAI Identifier: | oai:dadun.unav.edu:10171/22124 |
| Acceso en línea: | https://hdl.handle.net/10171/22124 |
| Access Level: | acceso abierto |
| Palabra clave: | Serum Sterols COPs Linearity Recovery Repeatability |
| Sumario: | A validated gas chromatography-mass spectrometry (GC-MS) detection method for the quantitative analysis of sterol oxidation products (SOPs) in serum is described. After a lipid extraction procedure with chloroform-methanol, a cold saponification and purification by solid phase extraction, oxysterols were derivatized to form trimethyl-sylil-ethers which were subjected to GC-MS analysis. Calibration curves for cholesterol oxidation products showed determination coefficient (R(2)) of 1.0, with low values for the coefficient of variation of the response factors (< 1%). Detection and quantification limits were below 5 ng/mL and 10 ng/mL, respectively. Recovery data were between 77.65% and 110.29% (CV < 10% for all compounds). Good results were obtained for within- and between-day repeatability, with values below 10%. In conclusion, the method performed is suitable for the determination and quantification of SOPs in serum. |
|---|