Biodegradable Zn−1wt.%Mg(−0.5wt.%Mn) Alloys: Influence of Solidification Microstructure on Their Corrosion Behavior

The development of biodegradable Zn-based alloys for implants that effectively mimic the functionality of native bone throughout the healing process is a multifaceted challenge; this is particularly evident in the task of achieving appropriate corrosion rates. This work explores the incorporation of...

Descripción completa

Detalles Bibliográficos
Autores: Vida, Talita, Cruz, Clarissa, Barros, André, Cheung, Noé, Brito, Crystopher [UNESP], Garcia, Amauri
Tipo de recurso: artículo
Estado:Versión publicada
Fecha de publicación:2023
País:Brasil
Institución:Universidade Estadual Paulista (UNESP)
Repositorio:Repositório Institucional da UNESP
Idioma:inglés
OAI Identifier:oai:repositorio.unesp.br:11449/308714
Acceso en línea:http://dx.doi.org/10.3390/surfaces6030019
https://hdl.handle.net/11449/308714
Access Level:acceso abierto
Palabra clave:corrosion
hardness
microstructure
solidification
Zn alloys
Descripción
Sumario:The development of biodegradable Zn-based alloys for implants that effectively mimic the functionality of native bone throughout the healing process is a multifaceted challenge; this is particularly evident in the task of achieving appropriate corrosion rates. This work explores the incorporation of 0.5wt.%Mn into a Zn−1wt.%Mg alloy, with focus on the relationship between corrosion behavior and microstructure. Electrochemical corrosion tests were carried out in a 0.06 M NaCl solution using as-solidified samples with two distinct microstructural length scales. Mn addition was found to induce significant electrochemical active behavior. Localized corrosion was predominant in interdendritic regions, with the ternary alloy exhibiting a higher susceptibility. For both alloys, the coarsening of the microstructure promoted a slight inclination to accelerate the corrosion rates in both biodegradable Zn alloys. The corrosion rate showed an increase of about nine-times with Mn addition for coarser eutectic spacings, while for finer ones, the increase was by about 22 times.